Error Messages
Error
Info text
(is displayed in case of fault when
the Info key is pressed)
Problem
Possible causes
ERR 99
DEVICE FAILURE
Error in factory settings
EEPROM or RAM defective
This error message only occurs in the
case of a total defect. The device must be
repaired and recalibrated at the factory.
ERR 98
CONFIGURATION ERROR
Error in configuration or
calibration data
Configuration or calibration data defec-
tive; completely reconfigure and recali-
brate the device.
ERR 95
SYSTEM ERROR
System error
Restart required.
If error still persists, send in the device
for repair.
ERR 01
NO SENSOR
Sensor error
Device type not assigned
Defective sensor
Sensor not connected
Break in sensor cable
ERR 02
WRONG SENSOR
Wrong sensor
ERR 04
SENSOR FAILURE
Failure in sensor
ERR 05
CAL DATA
Error in cal data
ERR 10
CONDUCTANCE TOO HIGH
Display range exceeded
Conductivity:
Conductance > 3500 mS
ERR 11
RANGE
Display range violation
ERR 13
TEMPERATURE RANGE
Temperature range violation