Diaphragm liquid pump NF 1.100 EX
Use
KNF Flodos AG BA_NF1.100EX_EN_05_152147
English translation of original Operating and Installation Instructions
5
2.4.4. Explosion Groups
Flammable gases and vapours are classified into explosion groups
(I, IIA, IIB and IIC) and temperature classes. Tab. 5 shows the
classifications of the most common flammable gases and vapours.
T1
T2
T3
T4
T5
T6
I
Methane
–
–
–
–
–
IIA Acetone
Ethane
Ethyl acetate
Ammonia
Ethyl chloride
Benzene
Acetic acid
Carbon monoxide
Methane
Methanol
Methyl chloride
Naphthalene
Phenol
Propane
Toluene
i-amyl acetate
n-butane
n-butyl alcohol
Cyclohexanone
1,2 dichloroethane
Acetic anhydride
Benzene
Diesel fuel
Aviation turbine
fuel
Heating oils
n-hexane
Acetaldehyde
–
–
IIB Town gas
Ethylene
Ethyl alcohol
Hydrogen
sulphide
Ethyl ether
–
–
IIC Hydrogen
Acetylene
–
–
–
Carbon
disulphide
Tab. 5
The classification of gases and vapours into groups according to
explosion group and temperature class applies to both the medium
to be transferred and the environment of the pump.
The pumps must only be used to transfer gases and vapours that
are either not explosive or belong to explosion groups II A or II B
and temperature class T4 (or lower) (shaded area in Tab. 5).
The pumps must only be used in an environment containing an
atmosphere that is either not explosive or belongs to explosion
groups II A or II B and temperature class T4 (or lower) (shaded
area in Tab. 5).
Transferred medium
Pump environment