7.5 Adjusting the flow rate
Set the pump to the required flow rate using the control knob.
The flow rate setting is shown by the flow rate indicator around the
control knob.
Type
Flow rate [l/min]
LIQUIPORT 100
0.2 to 1.3
LIQUIPORT 1.100
0.2 to 1.3
LIQUIPORT 300
0.5 to 3.0
LIQUIPORT 1.300
0.5 to 3.0
Tab. 15: Specified flow
If the analog input is active, the "EXT" indicator illuminates and the
control knob has no function. The flow rate is then specified by the
analog input. See Chapter 8.1
The flow rate of the pump is reduced by counterpressure, suction
head and higher medium viscosity.
In the event of counterpressure, the pump will not start at low flow
rate settings.
Reduce counterpressure or select a higher setting.
In the event of counterpressure the pump cannot be set to the lower
flow rates. The counterpressure must be reduced or a higher flow rate
must be selected.
Fig. 17: Operating range
Fig. 16: Adjusting the flow rate
Operation in the event of counter-
pressure