8.3. Dismantling the pump head
(FEM 1.02 / FEM 1.09 / UFEM 1.09)
Use a screwdriver to undo the head screws (
1
)
Remove the head plate (
2
), connecting plate (
3
) and intermediate plate (
4
)
Removing the valves and seals
Remove valve seats (
5
)
Remove seals (
6
), valve disks (
7
) and O-rings (
8
)
Cleaning the parts
Clean the diaphragm (
9
), valve disk (
7
), seals (
6
) and O-rings (
8
) with a
cloth and then blow off with compressed air
Blow off the intermediate plate (
4
) and the connecting plate (
3
) with com-
pressed air
Mounting the valves and seals
Mount O-rings (
8
) and valve disk (
7
) on valve seat (
5
)
Insert seal (
6
) in connecting plate (
3
)
Insert valve seats with valve disks (
7
) and O-rings (
8
) into connecting
plate (
3
) (slight pressure required)
Insert the seal (
6
) in the intermediate plate (
4
)
Mounting the pump head
Assemble the intermediate plate (
4
), connecting plate (
3
) and head plate
(
2
)
Push the assembled components onto the pump housing with finger
pressure
Secure the pump head to the housing using the 4 head screws (
1
)
The maximum tightening torque for the head screws is 1 Nm.
Head screws that have been undone before should be carefully
screwed in by hand so that the screws engage in the existing
threads.
Fig. 15: FEM 1.02 head cleaning
Fig. 16: FEM 1.09 head cleaning
1
Head screw
2
Head plate
3
Connecting plate
4
Intermediate plate
5
Valve seat
6
Seal
7
Valve disk
8
O-ring
9
Diaphragm