34
Troubleshooting
AZURA® Detector UVD 2.1S Instructions, V6820
7.2 Error messages
If other error messages are displayed besides those listed below, please
restart the device. Inform the Technical Support of the manufacturer in
case the error message displays repeatedly.
Display
Software
Measure
EXTERNAL
External error
Check the external devices and
cable connections.
CALIBR
Calibration failed
Restart the device. Check
whether lamps, motor and
filter are functioning cor-
rectly. Inform the Customer
Support in case the system
message repeats itself.
Restart calibration on the
device or in the chromato-
graphy software.
NO LAMP
D2 lamp operation
failed
Restart the device. If the error
occurs again, replace the lamp.
NO CALIB
Cannot operate an un-
calibrated instrument
Restart the device. Wait until
calibration is completed.
MOTOR
Motor failure
Restart the device. Inform the
Customer Support in case the
system message repeats itself.
WAVELEN
Wavelength stabiliza-
tion failed
Restart the device. Inform the
Customer Support in case the
system message repeats itself.
BUF.FULL
Data buffer overflow
Restart the device. Inform the
Customer Support in case the
system message repeats itself.
ZERO POS
Calibration failed – Ze-
ro order position not
found
Restart the device. Inform the
Customer Support in case the
system message repeats itself.