30
Operation
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700
5.3 Software operation
To operate the device with software, you have to establish a connection
between the LAN port and a computer.
5.3.1 Control methods
The device can be operated with
Mobile Control
OpenLAB® EZChrom Edition
ClarityChrom®
PurityChrom®
Chromeleon
TM
Note:
It is not possible to use 2 control methods simultaneously. If the
device is connected to the software, it cannot be controlled via Mobile
Control. The device status can however be monitored.
You find more information about supported software packages and firm-
ware versions in the list „KNAUER software device support“ (document
no. V1663):
www.knauer.net/softwarecontrol
You find a detailed description on chromatography software in the corres-
ponding software instructions.
Note:
3D data acquisition via RS-232 is not possible.
5.3.2 Operating with Mobile Control
The Mobile Control is a device control software which can be installed on
your computer or tablet. To control the device using the Mobile Control,
connect the computer or tablet with a Windows operating system to a
wireless LAN router.
Data transfer between device and Mobile Control is actualized through
wireless LAN. You find a detailed description on the Mobile Control in its
accompanying software instructions.
5.4 Meaning of the LEDs
There are three LEDs and a switch on the front of the device.
1
Left LED
2
Center LED
3
Right LED
4
Switch/Standby
button
1 2 3
4
Fig. 23
LEDs and switch/standby button on the front of the device