9
Technical Data
Designation
Data electrosurgical unit Minicutter
Mains connection
100–240 V; 50/60 Hz
Power consumption
without HF output
approx. 16 VA
at max. output power
approx. 120 VA
Protection class
I
IP classification
IP2X
Classification acc. to German
Medical Devices Act
II b
Leakage currents NF/HF
acc. to IEC 60601-2-2
Type
BF; defibrillator-proof
Nominal frequency
460/920/1,230 kHz
Modulation frequency
58/77 kHz
Operating mode
Intermittent INT 10 s / 30 s corr. to 25 % duty factor
Mains fuses
T 1.6 AH, 250 V
G 5 x 20 mm
Signal level
HF display: 55 dB(A)
Alarm: 65 dB(A)
Weight
approx. 3.9 kg
Radio interference suppression Limit values according to DIN EN 55011 and DIN EN 60601-2-2
Radio interference immunity according to IEC 801
Dimensions W × H × D
270 x 120 x 250 mm
Environmental conditions for
transport and storage
Ambient temperature
Relative humidity
Air pressure
-25
°C to +70
°C (270 °F to 273 °F)
10–100%
500–1,060 hPa
Environmental conditions for
operation
Ambient temperature
Relative humidity
Air pressure
+10
°C to +40
°C (270 °F to 273 °F)
30–75%
700–1,060 hPa
Compliant with 93/42/EEC