Electrosurgical Unit Minicutter
Revision 08
43
Bipolar coagulation
Coagulation current with high voltage for surface coagulation (fulguration).
Power setting range:
Max. 70 W at 50
Ω
Max. periodic HF output voltage:
Max. 340 V
max
Max. output current
1.200 mA at 50
Ω
Crest factor
2.5
Danger of injury from excessive voltage!
Active accessories to be used with this current must be able to withstand an HF voltage of at least 340 V.
It must be kept in mind that voltage is achieved even at settings below 10 and that the active accessory
must be selected accordingly.
Performance Diagram
H
F
po
we
r i
n
W
Setting