KLS Martin independant Care Vertical ES Скачать руководство пользователя страница 16

 

independant Care Vertical 

Instructions for Use 

16

 

Revision 01 

9

 

Carriage, Storage and Disposal 

In the following environmental conditions, carriage and storage conformity is needed 

 

Temperature: -10 °C to +40 °C (+14 °F to +104 °F) 

 

Humidity: 30 - 75% 

 

Atmospheric pressure: 700 - 1,060 hPa 

Do not overlap more than 5 devices and do not place them vertically. In case of irreparable damages or 
end-of-life equipment, the device must be disposed following the existing rules. If the user does not follow 
the rules above, he assumes full responsibility for creating potential effects to the environment and to 
human health. 

Below are shown the environment conditions’ use 

 

Temperature: -5 °C to +35 °C (+23 °F to +95 °F) 

 

Humidity: 30 - 75 % 

 

Atmospheric pressure: 700 - 1,060 hPa 

10

 

FAQ & Troubleshooting 

10.1

 

FAQ 

Question 

Answer 

Which are the materials independant Care is 

made of? 

Fixing structure (where provided) and screws – galvanized steel 
Frame, arms, pendant – epoxy painted aluminum 
Accessories – refer to the drawings (ABS, AVONITE, STEEL, STAINLESS 

STEEL, CDF, etc.) 

Is independant Care easy to install? 

The installation of independant Care is not particularly difficult if done 

following the instruction manual by a qualified staff 

Does independant Care ship to destination as 

separated elements? 

independant Care is shipped as a ready to install single component packed 

in cardboard box. Accessories are packed in cardboard box separately. 

Is independant Care easy to clean? 

Yes, independant Care is easy to clean due to smooth surfaces 

Is independant Care easy to maintain? 

No maintenance is needed except for gas implants (please refer to this 

manual) 

What does “declared carrying capacity” mean? 

The declared carrying capacity is to be intended as maximum payload as 

shown in drawings in section 6 “Functioning”, page 10 

Does independant Care have gas outlet prepiping 

included? 

It depends on request (please refer to QUOTATION and drawings) 

 

10.2

 

Troubleshooting 

Issue 

Cause 

Solution 

Electrical outlet power failure 

No power supply 

Check if power supply has been plugged correctly 

 

Содержание independant Care Vertical ES

Страница 1: ...Front Cover English independant Care Vertical ES independant Care Vertical MD independant Care Vertical LG Instructions for Use 0051 90 579 52 10 Revision 01 Date of Release 2020 12 www klsmartin com...

Страница 2: ...ould result in death or serious injury DANGER Indicates a situation which if not avoided will result in death or serious injury Instructions for Use Serial number Date of manufacture Manufacturer Dist...

Страница 3: ...plicable SOCKET 1 SOCKET 2 Socket identification as per electrical diagram where applicable xxxVA lighting Power consumption of each light Correct Configuration Key required Inlay shape diameter Click...

Страница 4: ...d Use 8 4 Configuration 9 5 Screw Torques 9 6 Functioning 10 7 General and technical features 14 7 1 General Features 14 7 2 Technical Specification 14 8 Cleaning and Maintenance 15 8 1 Recommended Ma...

Страница 5: ...o handle the device or product or use it for clinical applications please do not hesitate to contact the Product Management Phone 49 7461 706 345 Fax 49 7461 706 190 In case of technical questions and...

Страница 6: ...ion UDA installation must be done following indication listed on manufacturer installation manual Electrical connection must be done following the manufacturer electrical scheme Gas outlet distributio...

Страница 7: ...pplied on the plate Please don t connect adapters between the device s electrical sockets and plugs Risk of injury due to system parts falling down Please don t overload UDA s max load listed on the r...

Страница 8: ...ar other equipment 3 Intended Use UDA s pendant models independant Care are manufactured for medical use areas only in order to allow Electrical current flow Data nurse calling system phone PC network...

Страница 9: ...Care Vertical Instructions for Use Revision 01 9 4 Configuration independant Care Vertical ES independant Care Vertical MD independant Care Vertical LG 5 Screw Torques Dimension Torques M5 Nm 7 M6 Nm...

Страница 10: ...independant Care Vertical Instructions for Use 10 Revision 01 6 Functioning...

Страница 11: ...independant Care Vertical Instructions for Use Revision 01 11...

Страница 12: ...independant Care Vertical Instructions for Use 12 Revision 01...

Страница 13: ...independant Care Vertical Instructions for Use Revision 01 13...

Страница 14: ...n test For each unit the following tests are performed Ground impedance protection conform to standard 8 6 laid down by IEC EN 60601 1 Standard leakage current and dielectric strength measure conform...

Страница 15: ...Oglio BS Italy Disposal In compliance with provisions No 151 05 laid down by Legislative Decree Unit of Measures mm 8 Cleaning and Maintenance 8 1 Recommended Maintenance We recommend to check the fol...

Страница 16: ...nt epoxy painted aluminum Accessories refer to the drawings ABS AVONITE STEEL STAINLESS STEEL CDF etc Is independant Care easy to install The installation of independant Care is not particularly diffi...

Страница 17: ...elative humidity should be at least 30 Electrical fast transient 2 kV phase i Earth 1 kV phase i phase i 2 kV common mode 1 kV differential mode Voltage quality should be like a typical hospital envir...

Страница 18: ...n distance in meters Radio wave radiated power 3 V m 80 MHz to 2 GHz 3 V m P d 35 0 80 MHz to 800 MHz P d 7 0 800 MHz to 1 GHz P d 33 2 1 GHz to 2 5 GHz where P is the maximum output power rating of t...

Страница 19: ...12 Maintenance Log Book Serial number MD Official distributor Required maintenance Frequency Registration Name Date Electrical sockets functioning Fuse functioning Gas sockets functioning Lightning s...

Страница 20: ...sia Tel 7 499 792 76 19 russia klsmartin com KLS Martin Taiwan Ltd Taipei Taiwan Tel 886 2 2325 3169 taiwan klsmartin com Gebr der Martin GmbH Co KG Dubai United Arab Emirates Tel 971 4 454 16 55 midd...

Отзывы: