C-2
APPENDIX C Configuring and Updating the TPU
Series 5000 V2 Sonar System Operations and Maintenance Manual
P/N 11214512, Rev. 04
C.2 Querying or Changing the TPU IP Address
The TPU IP address is set up at the factory. The factory set default IP address is
192.168.0.81. However, should a different address be required, it can be changed.
To change the TPU IP address:
1.
Connect the supplied null modem cable to the NAV connector of the
TPU and to an available serial port on the SonarPro Workstation.
2.
Turn on the SonarPro Workstation and wait for the Windows desktop to appear.
3.
Turn on the TPU.
4.
Start Linux TPU Updater.
The
Klein Linux TPU Updater
dialog box opens.
5.
Click
Read TPU IP Address
.
The
Select RS-232 Port
dialog box shown in Figure C-2 opens.
6.
From the
Port
drop-down list box, select the serial port on the SonarPro
Workstation to which the TPU is connected.
7.
From the
Rate
drop-down list box, select the baud rate of the TPU’s NAV serial
port, or select
Determine Automatically
to have Linux TPU Updater find the rate.
The
Klein Linux TPU Updater
dialog box opens with the current IP address
displayed in the
TPU IP Address
text box as shown in Figure C-3.
Figure C-2:
The Select RS-232 Port Dialog Box.
Содержание 5000 V2 Series
Страница 24: ......
Страница 48: ......
Страница 62: ......
Страница 73: ...A 3 Figure A 1 Cable Length vs Towfish Depth Graph 1...
Страница 75: ...A 5 Figure A 3 Cable Length vs Towfish Depth Graph 3...
Страница 77: ...A 7 Figure A 5 Cable Length vs Towfish Depth Graph 5...
Страница 78: ......
Страница 82: ......
Страница 88: ......
Страница 92: ......
Страница 95: ...E 3...
Страница 97: ...E 5...
Страница 99: ...E 7...
Страница 101: ...E 9...
Страница 103: ...E 11...
Страница 105: ...E 13...
Страница 107: ...E 15...
Страница 109: ...E 17...
Страница 111: ...E 19...