-34
-
In the called-in notice in our example you can see the measuring range „0 ... 70
kN“. With the key which has now the effect of an arrow head
key you can call in further force-measuring ranges in kN.
The chosen force measuring range has to coincide with the measuring range of
the sensor.
Remark: The force measuring range is always stated on the decal of the sensor.
The picture on the left hand side explains the scope by means of the arrow head
key.
By pressing the key the choice is confirmed.
In order to be able to measure the pressure in the same way, the menu for the
pressure sensor must be called in by the key (please actuate the
key for 2 seconds).
For the user there are maximum six pressure measuring ranges in bar to choose
from.
The chosen pressure measuring range has to coincide with the measuring range
of the sensor.
Remark: The pressure measuring range is always stated on the decal plate of the
sensor.
The opposite picture explains the possibilities of choice by means of the arrow
head key.
By pressing the key the choice is confirmed.
2.
Preparations for force-sensing
resp. pressure measuring
2.2
Adjustment of firmly preset force measuring
ranges as calibration value
In order to measure a force, first of all the measuring range of the force sensor
has to be fed in.
The key has to be pressed for about 2 seconds.
* * * PROG P * * *
0 . . . 70.00 kN
(0 ... 15736 lbf)
p
0 . . . 40.0 kN
0 . . . 130.0 kN
Prog
ENT
F
Prog
ENT
* * * PROG P * * *
0 . . . 60.00 bar
(0 ... 870 psi)
0 . . . 200.0 bar
0 . . . 400.0 bar
0 . . . 600.0 bar
0 . . .1000.0 bar
2.3
Adjustment of firmly preset
pressure measuring ranges
as calibration value