27
EN
Emissions Test Standard
Compliance
RF emissions CISPR 11
Group 1, Class B
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Comply
Guidance and manufacturer’s declaration - electromagnetic Immunity
Radiated RF
IEC61000-4-3
(Test specification
for enclosure port
immunity to
RF wireless
communications
equipment)
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation
Modulation
(W)
Distance
(m)
Immunity
Test Level
(V/m)
385
380 - 390
TETRA 400
Pulse
modulation
18 Hz
1.8
0.3
27
450
380 - 390
GMRS 460,
FRS 460
FM
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704 - 787
LTE Band 13,17
Pulse
modulation
217 Hz
0.2
0.3
9
745
780
810
800 - 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
2
0.3
28
870
930
1720
1700
-
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1,3,4,
25; UMTS
Pulse
modulation
217 Hz
2
0.3
28
1845
1970
2450
2 400
-
2 570
BT,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
2
0.3
28
5240
5 100
-
5 800
WLAN 802.11
a/n
Pulse
modulation
217 Hz
0.2
0.3
9
5500
5785