2.3.17 Instantaneous Flow Zero Adjustment Function
The displayed value can be adjusted to "0" when the measured flow is
within ±5% F.S. of the zero point at the time of shipment from the factory.
<
Operation >
Press and button simultaneously over 3 sec. at the measurement
mode (not Accumulated flow value display mode) until display
[
00
]
.
And release holding the button to return measurement mode.
Measurement mode
Instantaneous flow
value return zero.
Press and button simultaneously
over 3 sec.
To release holding the button to
return measurement mode.