10
Alarms
10.1 Alarm functions
If a function of the unit is faulty or defective, an alarm is triggered.
Every alarm is displayed visually as well as acoustically.
The display alternates between the visual alarm and the tempera-
ture. The message is displayed until the alarm is acknowledged.
Acknowledging the alarm does not rectify the error.
The acoustic alarm is output as an alarm sound (hereafter referred
to as “buzzer”).
The unit is equipped with the following alarm functions:
n
Temperature alarm
n
Door-open alarm
n
Power failure warning
n
Alarm in case of defective display and control unit
Depending on the cause that triggered the alarm, the following
measures are required:
1 - Deactivate buzzer
“Deactivating the buzzer” on page 56
3 - Inform the service department
Tab. 9: Alarm functions (overview)
Alarm function
Display
Buzzer
Cause
Measure
Temperature alarm
3
n
The temperature is
over the tempera-
ture warning limit.
n
The remote warning
contact has been
triggered.
n
Deactivate buzzer.
n
Determine cause and rec-
tify.
3
n
The temperature is
below the tempera-
ture warning limit.
n
The remote warning
contact has been
triggered.
Door-open alarm
3
n
The door is open
for more than 60
seconds.
n
The door-open
alarm is not for-
warded via the
remote warning
contact.
n
Deactivate buzzer.
n
Close door.
Alarm functions
Alarms
Alarm functions
27.05.2021
Refrigerator/freezer MED series
55