Recommended temperature warning limits
The temperature warning limits must not be the
same as the target temperature.
Set the temperature limits as follows:
–
Upper temperature warning limit: at least 3 °C
higher than the target temperature
–
Lower temperature warning limit: at least 3 °C
lower than the target temperature
Recommended temperature warning limits for
FROSTER
To ensure that the unit functions properly, the tem-
perature warning limits must be set for FROSTER
Tab. 6: Temperature warning limits for MED and FROSTER MED
Model
Lower temperature
warning limit
Target temperature Upper temperature
warning limit
MED 100 PRO-ACTIVE
MED 126 PRO-ACTIVE
MED 200 PRO-ACTIVE
MED 288 PRO-ACTIVE
MED 340 PRO-ACTIVE
MED 468 PRO-ACTIVE
MED 520 PRO-ACTIVE
MED 520 CR PRO-ACTIVE
MED 600 PRO-ACTIVE
MED 720 PRO-ACTIVE
MED 288 ULTIMATE
MED 340 ULTIMATE
MED 468 ULTIMATE
MED 520 ULTIMATE
MED 520 CR ULTIMATE
MED 720 ULTIMATE
+2 °C
+5 °C
+8 °C
FROSTER MED 95 PRO-ACTIVE
-35 °C
-20 °C
-10 °C
Temperature warning limits
Commissioning
Programming the unit > Temperature warning limits
27.05.2021
Refrigerator/freezer MED series
36