15

 

 

7.

 

Technical specifications 

 

7.1.

 

Electrical 

 

Specifications 

 

Heater voltage 

12 ± 1 VDC 

Heater current 

460mA for the 7W version 

Heater power 

5.5W 

Ventilator voltage 

12 ± 1 VDC 

Ventilator current 

520mA ±20% 

Ventilator power 

6.3W 

Tacho signal 

5V and 475 Hz ±25% 

Dimensions 

210x178x126 mm 

Weight packed 

2 kg

 

Weigh unpacked 

1.6 kg

 

Ingression protection 

IP68 

Max cable length using Kipp & 
Zonen 8 wire cable 

25

 

Temperature range 

-40°C to +70°C 

Increase of air temperature 

0.25 K without heating 
1 K at 6.8 W heating 

Compliance with standards 

ISO/TR 9901, BSRN operations manual 

Zero o set A reduction 

50% reduction for CMP’s

 

Cable resistance 

0.16 

Ω

/m (2 x 0.08 

Ω

/m) 

Voltage drop at maximum 
heating 

0.08 V/m 

 

 

Intended instruments 

CMP6, CMP10, CMP11, CMP21, CMP22 
SMP6, SMP10, SMP11, SMP21, SMP22 
CMP3, SMP3 and CGR3 with reduced 
specifications 
ISO/TR 9901, BSRN operations manual 

 
Note: The performance specifications quoted are worst-case and/or maximum values 
 

7.2.

 

Dimensions and weight 

 

 

Weight without cable: 1.6 kg  

 

 

Содержание CVF4

Страница 1: ...Instruction Manual CVF4 Ventilation Unit V2007 01...

Страница 2: ...he product has been installed and used according to the instructions supplied by Kipp Zonen Kipp Zonen shall in no event be liable for incidental or consequential damages including without limitation...

Страница 3: ...8 4 5 Electrical connections 10 5 Accessories 11 5 1 Spare filters 11 5 2 CMF4 Mounting Fixture 11 5 3 CMB1 Mounting Bracket 11 5 4 Adapter for use of the CVF4 in the CM121C shadow ring 12 6 Maintena...

Страница 4: ...eters However with the CMP3 and SMP3 and CGR3 the ventilation effect will be less due to the larger opening around the dome The CVF4 is meant to run continuously The heater can either be switched on p...

Страница 5: ...his applies both to the product itself and to any accessories marked with the same symbol Disposal of any such item as unsorted domestic waste is not allowed 2 3 Customer support In case you need supp...

Страница 6: ...g holes CMP3 SMP3 and CGR3 4 Mounting holes all other instruments 5 Connector 6 Cover nuts 7 Slit for radiometer cable 3 2 CVF4 bottom side 1 Filter cover 2 Connector box 3 Mounting feet for SOLYS2 Fe...

Страница 7: ...washers and Allen key 4 Instruction sheet 5 Pack of 5 spare ventilation fan inlet filters For optional accessories see chapter 3 4 2 Tools required If the CVF4 is mounted with the supplied screws onl...

Страница 8: ...ure of the CVF4 on the SOLYS2 Sun Tracker on the next page The locations of the CVF4 mounting holes on the SOLYS2 top mounting plate are shown below CVF4 mounting holes When using your own mounting pl...

Страница 9: ...and radiometer levelling might need re adjustment After all mounting screws have been secured the cables of the CVF4 and the radiometer can be connected The radiometer cable goes down through the sli...

Страница 10: ...nal It can be connected to a data logger to remotely check proper operation of the CVF4 Ventilator and heater ground are separated but may be connected Ventilator and tacho ground are connected via a...

Страница 11: ...mber 2682047 from Kipp Zonen or under EBM Papst part number FP80T from a local distributor of Pabst fans 5 2 CMF4 Mounting Fixture Article number 0362703 5 3 CMB1 Mounting Bracket Article number 03697...

Страница 12: ...Adapter for use of the CVF4 in the CM121C shadow ring CVF4 s Remove CVF4s feet and Remount two feet in these counterbore holes Relocation of 2 of the CVF4s mounting feet to fit it on the CM121C shado...

Страница 13: ...adapter part is included with the CM121C Note CM121B Article number 0346900 is not supplied with nor aligned for the CVF4 adapter Article number 0346700 CM121C Article number 0346901 is aligned for a...

Страница 14: ...output is logged with the radiometer data the performance of the ventilator can be remotely checked 1 Refer to radiometer instruction sheet for maintenance instructions 2 Regularly inspect the fan inl...

Страница 15: ...le 25 m Temperature range 40 C to 70 C Increase of air temperature 0 25 K without heating 1 K at 6 8 W heating Compliance with standards ISO TR 9901 BSRN operations manual Zero o set A reduction 50 re...

Страница 16: ...en product please contact your local representative in the first instance The information can be found in the Contact section of our website at www kippzonen com 1 Alternatively you can contact us dir...

Страница 17: ...17...

Отзывы: