Kipp & Zonen CNR 4 Скачать руководство пользователя страница 32

 
 
 
 

 
 

31 

The CNR 4 body and ambient air should be at the same temperature as much as possible. Let the pyrgeometer rest 
for at least five minutes to regain its thermal equilibrium. Set the voltmeter to its most sensitive range. To test if the 
pyrgeometer is working properly, we suggest putting your hand in front of the pyrgeometer. The thermal radiation 
will  cause  pyrgeometer  to  generate  a  positive  voltage  when  the  hand's  surface  temperature  is  higher  than  the 
pyrgeometer  temperature.  The  pyrgeometer  will  generate  a  negative  voltage  if  the  hand  is  colder.  The  signal  is 
proportional to the temperature difference (see the rule of thumb of 1.1.6.5). The radiation that is emitted by the 
hand can be calculated by dividing the pyrgeometer output by the calibration factor, and subsequently correcting 
for the temperature, according to equation 1.2. Still no faults found? Your pyrgeometer is probably doing fine. 
 
 

5.3

 

Testing the Pt-100 

 

Using  a  meter,  which  measures  resistance,  you  can  check  the  operation  of  the  Pt-100.  If  connected  properly,  the 
resistance  of  two  opposite  wires  of  the  Pt-100  should  be  measured.  The  value  can  be  read  in  the  table  1.1  and 
should be above 100 Ohms (cable resistance should measure about 0.1 ohms per metre cable). When in doubt the 
thermistor resistance (temperature) can be checked as well for reference 
 

 

5.4

 

Testing the thermistor 

 
Using a meter, which measures resistance, you can check the operation of the thermistor. If connected properly, the 
resistance of two wires of the thermistor should be measured. The value can be read in the table 1.2 and should be 
around 10.000 Ohms for 25 °C. (cable resistance should measure about 0.1 ohms per metre cable). When in doubt 
the Pt-100 resistance (temperature) can be checked as well for reference 
 
 

5.5

 

Testing the Heater 

 

The optional CNF 4 consists of a heater and ventilator. Using a meter, which measures resistance you can check the 
operation of the heating resistor. The value should be around 15 Ohm. (Cable resistance should measure about 0.1 
ohms per metre cable). 
 
 
Using  a  meter,  which  measures  resistance,  you  can  check  the  operation  of  the  heater.  If  connected  properly,  the 
resistance of two opposite wires of the heater should measure about 8 ohms (this includes the cable resistance for 
the  standard  10-meter  cable).  The  cable  resistance  should  measure  about  0.1  ohms  per  metre  cable.  An  infinite 
resistance reading indicates the likelihood of a broken wire, or cable. 
 
 

5.6

 

Testing the Ventilator 

 

The impedance of the ventilator motor can be checked to tested for reference The value should be around 30 Ohms 
(cable  resistance  should  measure  about  0.1  ohms  per  metre  cable).  In  this  case  a  correct  value  is  measured  this 
does not guarantee proper operation. It is possible the ventilator is stalled by an object blocking the fan. This can 
be checked by removing the cover and filter and inspecting the rotation of the fan by hand. 
 

Содержание CNR 4

Страница 1: ...Net Radiometer CNR 4 Instruction Manual...

Страница 2: ...LITY Kipp Zonen guarantees that the product delivered has been thoroughly tested to ensure that it meets its published specifications The warranty included in the conditions of delivery is valid only...

Страница 3: ...nder our sole responsibility that the product Type CNR 4 Name Net Radiometer To which this declaration relates is in conformity with the following standards Imissions EN 50082 1 Group standard Emissio...

Страница 4: ...f the pyranometer 18 2 2 2 Spectral properties of the pyranometer 20 2 2 3 Directional Cosine response of the pyranometer 20 2 3 Properties of the pyrgeometer 22 2 3 1 Window heating offset 22 2 3 2 W...

Страница 5: ...ient air flow over the domes and windows to minimize the formation of dew and reduce the frequency of cleaning The integrated heater can be used to melt frost The CNR 4 specifications when used with C...

Страница 6: ...as also 8 pins but only 6 wires are coming out of the cable Exchanging the connectors will not damage the CNR 4 To prevent mix up of the cables after installation it is advised to mark the cables with...

Страница 7: ...rgeometer Upper Oben Sup rieur Superior Pyrgeometer Lower Unten Inf rieur M s bajo Blue Blau Bleu Azul Connect with Anschluss an Relier Conectar con Yellow Gelb Jaune Amarillo Green Gr n Vert Verde Ho...

Страница 8: ...d covered with a black sleeve If the CNR 4 rod is grounded to the mast it is advised not to connect the shielding on the data logger side This might cause ground loops and offsets in the signal Figure...

Страница 9: ...ween the pyrgeometer and the object that it is facing This implies that the pyrgeometer will generate a positive voltage output V when it faces an object that is hotter than its own sensor housing and...

Страница 10: ...in a positive signal To measure the Far Infrared irradiances with the two pyrgeometers separately the Pt 100 output is required The formula 1 2 is used to calculate the Far Infrared irradiance of the...

Страница 11: ...39 100 78 101 17 101 56 101 95 102 34 102 73 103 12 103 51 103 90 104 29 104 68 105 07 105 46 105 85 106 24 106 63 107 02 107 40 107 79 108 18 108 57 108 96 109 35 109 73 110 12 110 51 110 90 111 28...

Страница 12: ...2952 2854 Table 1 2 Resistance values versus temperature in C of the CNR 4 s thermistor Relatively small errors occur when the CNR 4 is not in thermal equilibrium This happens for example when the hea...

Страница 13: ...pyranometer E lower pyranometer 1 4 In this formula E is calculated according to formula 1 1 Net Solar radiation will always be positive Checking this can be used as a tool for quality assurance of y...

Страница 14: ...ature column nine tracks the ambient temperature Under cloudy conditions this is logical cloud bases will be colder than the ambient temperature at instrument level the temperature difference depends...

Страница 15: ...e upper pyrgeometer Figure 1 2 3 clear day for the downward facing pyrgeometer Upper pyrgeometer day with alterna3ng cloud elds pyrgeometer U_emf Sensi3vity W m Temp YSI 44031 C Upwelling signal downw...

Страница 16: ...cooling The sky and the ground temperature can be calculated from the measured values of the sensors using formulas 1 6 and 1 7 day night Cloudy clear 20 C 20 C pyrgeo Up pyrgeo low pyrano up pyrano...

Страница 17: ...g improve the accuracy significantly Due to the fact that the Net Solar radiation can be very intense 1000 Watts per square metre compared to a typical 100 for the Net Far Infrared radiation the accur...

Страница 18: ...ature C 40 to 80 CE according to EC guideline 89 336 EEC 73 23 EEC Environmental protection IP 67 2 2 Properties of the pyranometer The pyranometer consists of a thermopile sensor and a glass dome bot...

Страница 19: ...sensitivity N S 40 C to 10 C 4 10 C to 40 C N S 40 C to 80 C Definition Maximum error due to any change of instrument temperature within the indicated temperature interval Tilt error 1 Definition Max...

Страница 20: ...nition Deviation of the product of spectral absorption and spectral transmittance from the corresponding mean within the indicated spectral range Uncertainty in daily total 5 95 confidence level Defin...

Страница 21: ...t absorbs all radiation from all wavelengths Its field of view is 180 degrees The directional properties are similar to that of a blackbody Another way of expressing these directional properties is to...

Страница 22: ...sponse or cosine response of the pyranometer On the horizontal axis the zenith angle is shown 0E zenith angle equals 90E angle of incidence The vertical axis shows the deviation from the ideal cosine...

Страница 23: ...range This error is neglected however in the net radiation calculation this is justified because the solar radiation is always dominant when this error occurs Due to its construction the window heatin...

Страница 24: ...indow heating offset W m 6 0 to 1000 W m solar irradiance Definition Caused by heating of the dome due to solar radiation Field of view Upper detector Lower detector 180 150 Definition Sensor opening...

Страница 25: ...ectivity of the total instrument 2 3 5 Directional Cosine response of the pyrgeometer The measurement of the Far Infrared radiation falling on a surface also called irradiance or radiative flux requir...

Страница 26: ...e significant Dew deposition will probably take place under conditions with large Far Infrared irradiation from the pyrgeometer detector to the clear sky typically minus 100 Watts per metre square Und...

Страница 27: ...A table stating the resistance values can be found in table 1 1 The thermistor has a much larger resistance value than the Pt 100 also the change in resistance with respect to temperature in absolute...

Страница 28: ...librated against the World Reference at WMO in Davos 3 3 Recalibration of pyranometers and pyrgeometers We suggest recalibration for all sensors pyranometers and pyrgeometers to be performed every two...

Страница 29: ...nt by less than 1 The Net Radiometer should be installed horizontally using the level on the body of the CNR 4 For installation in buildings or in solar energy applications you will often have to moun...

Страница 30: ...or cover black part in picture below can be removed by just pulling it down from the CNR 4 The filter needs to be checked for dust and particles every 6 12 months It can be cleaned by simply washing i...

Страница 31: ...e starting the second test measurement let the pyranometer rest for at least five minutes to let it regain its thermal equilibrium For testing set a voltmeter to its most sensitive range setting Darke...

Страница 32: ...ermistor Using a meter which measures resistance you can check the operation of the thermistor If connected properly the resistance of two wires of the thermistor should be measured The value can be r...

Страница 33: ...bject The top U bolts allow rotation of the CNR 4 rod while the extra screw under the front U bolt allows the rod to tilt All mounting material for fixation to a horizontal or vertical pole is include...

Страница 34: ...33 Fig 7 2 CMB 1 mounting examples with CNR 4 albedometer and pyranometer...

Страница 35: ...clude the following items A delivery includes One CNR 4 One mounting rod Two cables Two drying cartridges Four calibration values on 2 calibration sheets long and short wave One CD with instruction ma...

Страница 36: ...Here recalibration to the highest standards can be performed at low cost Recalibration can usually be performed within four weeks If required urgent recalibration can be accomplished in three weeks or...

Страница 37: ...ro de repuestos Passion for Precision Go to www kippzonen com for your local distributor or contact your local sales office HEAD OFFICE Kipp Zonen B V Delftechpark 36 2628 XH Delft P O Box 507 2600 AM...

Отзывы: