INSTALLATION
29
3.3 ELECTRICAL
CONNECTION
The CM 11 is provided with a 10 m cable with three leads and a shield
covered with a black sleeve.
The colour code is:
red
=
plus
blue
=
minus
white
=
case
The shield is isolated from the case, so no shield-current can exist. Shield and
white lead may be connected to the same ground at the readout equipment.
The cable must be firmly secured to minimise spurious response during
stormy weather (pressing the standard cable produces voltage spikes, a tribo
electric effect and capacitance effect).
Kipp & Zonen pyranometer cables are of low noise type, however take care
that the terminals '+' and '-' at a connection box have the same temperature,
to prevent thermal EMF's. A box or connector with metal outer case is
advised.
Looking at the circuit diagram of figure 14, it is clear that the impedance of the
readout equipment is loading the thermistor circuit and the thermopiles. It can
increase the temperature dependency of the pyranometer. The sensitivity is
affected more than 1% when the load resistance is under 150 k
Ω
. For this
reason we recommend the use of readout equipment with input impedance’s
of 1 M
Ω
or more such as potentiometric recorders, digital voltmeters, etc. The
solar integrators and chart recorders from Kipp & Zonen meet these
requirements. Long cables may be used, but the cable resistance must be
smaller than 0.1% of the impedance of the readout equipment.
Kipp & Zonen supplies shielded low-noise extension cable up to lengths of
200 m. This extra length can be supplied fitted to the pyranometer when
ordered, or can be coupled by waterproof connectors to the CM 11 cable. The
lead resistance is 8 Ohm/100 m.
Содержание CM 11
Страница 1: ...INSTRUCTION MANUAL Kipp Zonen 0305 201 PYRANOMETER ALBEDOMETER 14 11...
Страница 13: ...GENERAL INFORMATION 12...
Страница 17: ...TECHNICAL DATA 16 Figure 5 Kipp Zonen pyranometer CM 11 outline dimensions in mm...
Страница 18: ...TECHNICAL DATA 17 Figure 6 Kipp Zonen albedometer CM 14 outline dimensions in mm...
Страница 25: ...INSTALLATION 24 Figure 10 Mounting the pyranometer Screw M5x80 Nylon insulator Washer M5 Nut M5...
Страница 35: ...OPERATION 34...
Страница 43: ...CALIBRATION 42...
Страница 47: ...TROUBLE SHOOTING 46...
Страница 51: ...APPENDIX I 50...
Страница 53: ...APPENDIX II 52...
Страница 55: ...APPENDIX IV 54...
Страница 57: ...APPENDIX IV 56...
Страница 59: ...APPENDIX IV 58...
Страница 61: ...APPENDIX IV 60...
Страница 63: ...APPENDIX IV 62...