PART VII    AFTER-SALE SERVICE  

Please observe the correct operation of this biomedical freezer. If there is any issue, please troubleshoot the error with the following table. If the 
issue cannot be fixed, please inform our service center. We will serve you as quickly as possible to avoid any loss items or time.  

AFTER-SALE SERVICE

BIOMEDICAL FREEZERS

PROBLEM

POSSIBLE CAUSE

Freezer is not running

Power outage has occurred

 

The plug is bad or not securely plugged-in

 

Fuse is blown

 

Voltage is too low or high

Compressor is not running

Power in control panel is off

 

Temperature setting is wrong

The temperature does not reach the set value 

The door is not closed properly or is opened too frequently

 

There is too many items in the unit

 

Outside temperature is too high

Loud noise

The unit is not level

 

The unit is leaning on a wall

Surface condensation on door

When the weather is humid, condensation might collect on the door. This can be expected. Wipe 
condensation with dry cloth.

The door is not closed properly, and the cool air 

leaks 

After the freezer is used for a period of time, the door seal becomes hard and deformed. 
Maintenance method: Blow the deformed part of the hot seal with a blower to soften it, and then 
close and compress it after the door seal becomes soft. 

Alarm flashing, buzzing

If you just put in an item, the temperature will need to stable and the alarm will reset automatically 
after running for a period of time.

 

If the door is not closed tightly, the door open alarm will sound.

 

If the battery is low, the alarm will flash and will turn off after a period of time. 

 

Alarm will flash when the temperature exceeds the set temperature.

The below may happen during normal operation: 

There are some light noises when the compressor starts up and stops. 

After opening the door and placing in hot items, the control system will show high temperature and/or high humidity alarm.  

Solution

: hot items should be cooled by natural cooling before placing in unit. Do not put too many items in at one time. After the temperature 

is stabilized, the high temperature and/or high humidity alarm will turn off.  

You will hear the sound of running water through the refrigerant pipes.  

Before calling the service engineer, please clean and disinfect the refrigerator.   

Condition: do not shake heavily or strike. Avoid rain or other water drenching the unit.  

Storing environment temperature:-40°C~+55°C, Relative humidity: 10%~ 90%.

Содержание MLD450

Страница 1: ...MEDICAL FREEZERS BEFORE USE PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS Applicable Models MLF450 MLD450 MLRF450 MLF450 Biomedical Freezer MLD450 Ultra Low Temperature Freezer ML...

Страница 2: ...2...

Страница 3: ...t be held responsible for any adverse consequences However the user can apply for verification of third party accessories from KingsBottle before use You should inspect and maintain the device accordi...

Страница 4: ...the unit do not push the power line up against the wall or other furnishings Leave room for the compressor and heat source as well Use the dedicated power supply according to the instructions on the...

Страница 5: ...items that are not suitable for low temperature settings During usage the temperature inside the unit is very low Do not touch the interior surface or any objects inside without protective gear Always...

Страница 6: ...the displayed temperature and the actual temperature in the center of the unit at first The center temperature will gradually approach the displayed temperature To clean the interior surface use luke...

Страница 7: ...nnected to circuit protection facilities Check the working voltage of the place before starting the freezer A voltage stabilizer is suggested to be used at the place where the voltage is not stable Ma...

Страница 8: ...of refrigeration there is a little difference between the actual temperature displayed on the controller and the set temperature This is a normal phenomenon Biomedical freezers are storage devices do...

Страница 9: ...ion piece Structure and composition the product consists of a box a glass door structure a refrigeration system and a control system Scope of application This product is suitable for hospitals pharmac...

Страница 10: ...ed by pressing the up key for 3 seconds g is the down key In the parameter setting mode this will move to the previous parameter or reduce the parameter value When setting the set temperature the temp...

Страница 11: ...cked at this time After unlocking press for 3 seconds the digital display will show the parameter code PS1 enter the desired parameters Use to adjust Display PS1 b1 b2 Set1 H01 L01 Set2 H02 L02 n y r...

Страница 12: ...mperature is too high the high temperature alarm shows H2 8 L02 0 0 10 0 5 0 Low temperature alarm set value L L 0 when the alarm is cancelled When the temperature is too low the low temperature alarm...

Страница 13: ...the cooling process happen for as long as possible before opening the door This will cause the temperature to rise Due to the location of the thermostat the unit might have slightly different temperat...

Страница 14: ...freezer should be cleaned and maintained For the sake of safety first unplug the unit Then wipe the inner and outer surfaces of the freezer with a soft cloth CAUTION DO NOT sprinkle water on the surf...

Страница 15: ...hard and deformed Maintenance method Blow the deformed part of the hot seal with a blower to soften it and then close and compress it after the door seal becomes soft Alarm flashing buzzing If you jus...

Страница 16: ...e to the negative pole c Remove the two screws from the battery plate with screwdriver then remove the battery d New replaced battery model BT 12M4 0AC 12V4 0AH e The replacement battery is recyclable...

Страница 17: ...minum plate with spraying Aluminum plate with spraying Shelves 6 2 ABS 6 2 ABS R 3 coated steel wired shelf F 6 ABS External Lock Yes Yes Yes Lighting LED LED LED Access Port 2pcs 25 mm 2pcs 25 mm 2pc...

Страница 18: ...authorized repair or any other cause beyond the control of the seller whether similar or dissimilar to the foregoing Purchaser understands and acknowledges that the goods sold here are Biomedical Free...

Отзывы: