PART V DISPLAY PANEL FUNCTION DESCRIPTION
OPERATION
1. Function description of control panel
a.
The temperature display will show the average
temperature of the upper chamber and the lower chamber inside the
unit while under normal conditions. The temperature is displayed in
Celsius.
b.
Door status indicator
When the door opened, the light is on; when the door is closed, the light
is off.
c. Fault indicator light: When the product runs normally, the indicator
light is off; The indicator light is on for operation abnormality.
d.
is the button to turn on the light.
The default state of the lamp is off. Use this key to turn the light on
and off. When off, the light will turn on when the door is open and off
when closed. If turned on, the light will remain on with the door is
opened or closed. The unit has two settings: automatic lighting and
manual lighting. With automatic lighting, the light will turn on when the
door is opened and will turn off after 5 seconds. The user can
manually control the light with this button. When the light is manually
turned off, the unit will enter the automatic lighting state.
e.
is the printing key.
The system can keep the data for 7 days to print. Press the print key
to print the temperature at any time.
f.
is the up key.
In the parameter setting mode, you can move to the next parameter,
or increase the parameter value. For example, when setting the set
temperature, this will increase the temperature. When setting
parameters, the up button will increase the amount of each
parameter. In normal state, data can be imported by pressing the up
key for 3 seconds.
g.
is the down key.
In the parameter setting mode, this will move to the previous
parameter or reduce the parameter value.
When setting the set temperature, the temperature is reduced when
pressed.
h.
is the set / mute key. Press this for the inside temperature.
After 5 seconds, the normal display is restored.
While the unit is unlocked, press
for 3 seconds to enter the
unit’s menu.
When the alarm is sounding (including the cabinet high temperature
alarm, the door open alarm, the sensor fault alarm, etc.), press
and the buzzer will stop. The display will show the temperature for 5
seconds before returning to normal display.
While the unit is unlocked, you can use
as the settings key.
In the unlocked state, when setting parameter mode, press this key
to display the parameter and parameter name. If the press time is
longer than 3 seconds, the settings will be saved and the display will
return to the normal interface.
i. USB data export.
Automatic export:
when the USB disk is connected to the USB
interface, the unit will buzz once, display "on", and the PDF files are
generated for the current month and the last month. After the data
transmission is completed, the unit will buzz once again, display
"end", and then will return to normal display after 6 seconds
Manually export:
once the USB disk is inserted, press the
for
3 seconds. The display will show “d01” and you can press up or
down to select from d01-d012: previous one month to previous
twelve months. Select the number of months then press
to
confirm the data you want downloaded.
Note:
when the digital display flickers "LoF", the recorder has not
started; press
at the same time, the "LoF" will disappear
after 3 seconds ,the buzzer will sound once, and the recorder will
start.