MAINTENANCE
This Initialisation Procedure is concerning the lifting mechanism
on the following products:
Kingkraft Variable Height Easibath
If the bath will not respond to the pressing of the raise or lower buttons on the handset /
membrane switch, and you can hear a beeping sound coming from the bath, you will have to
follow this initialisation sequence to re-set the bath.
Failure of the mains power to the bath, or running the bath off the back-up battery only could
lead to the synchronisation of the lifting columns becoming compromised, causing the system
to stop working until it is reset.
Initialisation Procedure