Kingfisher KI6106 Series Скачать руководство пользователя страница 4

 

Revision 2                                             Date: 11 Jun 2020                                                                                      Page 4 of 12 

2.  Technical specifications 

 

 

 

Standard wavelengths 

(nm) 

1270 

(upstream) 

1310 

(upstream) 

1490 

(downstream) 

1577 

(downstream) 

Pass zone(nm) 

1260 

 1280 

1290 

 1330 

1470 

 1505 

1570 ~1585 

Range(dBm) 

-30 

 10 

-30 

 10 

-45 

 10 

-45 

 10 

Isolation @1270 (dB) 

 

> 30 

> 40 

> 40 

Isolation @1310 (dB) 

> 30 

> 40 

> 40 

Isolation @1490 (dB) 

> 40 

> 40 

> 40 

Isolation @1577 (dB) 

> 40 

> 40 

> 40 

Uncertainty(dB) 

± 0.5 

Polarization Dependent 
Loss (dB) 

< 0.25 

Linearity(dB) 

± 0.1 

Insertion Loss(dB) 

< 1.5 

Resolution 

0.01dB 

Measurement unit 

dBm / dB / W / Pass / Fail 

Memory capacity 

99 records of 4

λ

with date/time stamp  

Auto power off time 

10 minutes after last key press 

Battery type 

7.4V 1000mAH rechargeable Lithium

 battery 

Battery life 

10 hours 

Power supply/charger 

Input 

100~240V, 50/60Hz 0.3 A

; Output: 

8.4V, 0.5A

  

Connector: 

1.3(ID) x 3.5(OD) x 9.5 (L) mm

 

ID=positive, OD=negative

 

Operating / storage 
temperature 

-10 ~ 50 

 / -25 ~ 70 

 

Dimensions / Weight   

200 * 90 * 43  mm /   330g 

Содержание KI6106 Series

Страница 1: ...ed repair of the equipment This Kingfisher International product is guaranteed against defective components and workmanship for a period of 1 year from the date of delivery unless specifically stated...

Страница 2: ...ON OFF instrument 5 3 2 Auto Power Off mode 5 3 3 Setting Time 5 3 4 Selecting measurement unit 6 3 5 Relative Measurement Mode 6 3 6 Alert Pass Fail threshold setup 7 3 7 Measurement with Pass fail D...

Страница 3: ...2 1 The instrument Rubber holster Captive dust cap Battery door Product label LCD Keypads Charging jack USB connector Rubber flap Optical connector ports ONT left for upstream 1270 1310 nm tests OLT r...

Страница 4: ...40 40 Uncertainty dB 0 5 Polarization Dependent Loss dB 0 25 Linearity dB 0 1 Insertion Loss dB 1 5 Resolution 0 01dB Measurement unit dBm dB W Pass Fail Memory capacity 99 records of 4 with date time...

Страница 5: ...Off mode is enabled instrument will switch off automatically if no subsequent key is pressed 10 minutes after it was switched on To disable enable Auto Power Off mode press OFF in turn 3 3 Setting Ti...

Страница 6: ...and time format Date YY MM DD Time HH MM Time Setting mode will exit automatically if no subsequent key is pressed 10 sec after it was selected 3 4 Selecting measurement unit At home display press F2...

Страница 7: ...ve or negative of the value After all the 3 threshold values see below for the selected wavelength are correctly set Alert displayed as Pass displayed as Fail displayed as x The alert pass fail thresh...

Страница 8: ...urn to home display 3 7 Measurement with Pass fail Display Mode When performing measurements in this mode measurement results based on the threshold values set in section 3 6 for each wavelength will...

Страница 9: ...the above Data Saving mode will exit automatically if no subsequent key is pressed 10 sec after being selected To retrieve saved measurement data Press and hold to enter Data Browsing mode indicate by...

Страница 10: ...rged Standard instrument comes with a charger of US type AC plug If require order suitable AC plug adapter from Optional Accessories section below 4 Maintenance Keep the sensors surfaces clean and fre...

Страница 11: ...emove the batteries and dry it out carefully for at least one hour before using it again Storage During prolonged storage remove batteries to eliminate the possibility of acid leakage Use only high qu...

Страница 12: ...Revision 2 Date 11 Jun 2020 Page 12 of 12 7 Optional Accessories Description P N AC Power Adapter Plug US to UK OPT093 AC Power Adapter Plug US to AUS OPT094 AC Power Adapter Plug US to EUR OPT095...

Отзывы: