INTRODUCTION

INTRODUCTION

4

5

CAUTION

* Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a 

medicine prescribed by a doctor. Consult your doctor if you have any question about your blood pressure.

* When the device was used to measure patients who have common arrhythmias such as atrial or 

ventricular premature beats or atrial fibrillation, the best result may occur with deviation. Please consult 

your physician about the result.

* Don't kink the connection tube during use, otherwise, the cuff pressure may continuously increase which 

can prevent blood flow and result in harmful injury to the PATIENT.

* When using this device, please pay attention to the following situation which may interrupt blood flow and 

influence blood circulation of the patient, thus cause harmful injury to the patient: connection tubing kinking 

too frequent and consecutive multiple measurements; the application of the cuff and its pressurization on 

any arm where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present; inflating the 

cuff on the side of a mastectomy.

* Warning: Do not apply the cuff over a wound;otherwise it can cause further injury.

*Do not inflate the cuff on the same limb which other monitoring ME equipment is applied around 

simultaneously, because this could cause temporary loss of function of those simultaneously-used 

monitoring ME equipment.

*On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff 

immediately. Prolonged high pressure (cuff pressure 

 300mmHg or constant pressure 

 15mmHg for 

more than 3 minutes) applied to the arm may lead to an ecchymosis.

*Please check that operation of the device does not result in prolonged impairment of patient blood 

circulation.

* When measurement, please avoid compression or restriction of the connection tubing.

* The device cannot be used with HF surgical equipment at the same time.

* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically 

investigated according to the requirements of ISO 81060-2:2013.

* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the manufacturer.

* This device is contraindicated for any female who may be suspected of, or is pregnant. Besides providing 

inaccurate readings, the effects of this device on the fetus are unknown.

* Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries.

* This unit is not suitable for continuous monitoring during medical emergencies or operations.Otherwise, 

the patient’s arm and fingers will become anaesthetic, swollen and even purple due to a lack of blood.

* When not in use, store the device in a dry room and protect it against extreme moisture, heat, lint, dust 

and direct sunlight. Never place any heavy objects on the storage case.

* This device may be used only for the purpose described in this booklet. The manufacturer cannot be held 

liable for damage caused by incorrect application.

*This device comprises sensitive components and must be treated with caution. Observe the storage and 

operating conditions described in this booklet.

* The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable 

anesthetic mixture with air of with oxygen or nitrous oxide.

* Warning: No servicing/maintenance while the ME equipment is in use.

* The patient is an intended operator.

* The patient can measure data and change battery under normal circumstances and maintain the device 

and its accessories according to the user manual.

* To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated 

interference signal or electrical fast transient/burst signal.

* The blood pressure monitor and the cuff are suitable for use within the patient environment. If 

you are allergic to polyester, nylon or plastic, please don't use this device. 

* During use, the patient will be in contact with the cuff. The materials of the cuff have been tested and 

found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any 

potential sensization or irritation reaction.

CAUTION

* This device is intended for adult use in homes only. 

* The device is not suitable for use on neonatal patients, pregnant women,patients with implanted, 

electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, 

peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or 

people who received a mastectomy. Please consult your doctor prior to using the unit if you suffer from 

illnesses.

* The device is not suitable for measuring the blood pressure of children. Ask your doctor before using it 

on older children.

* The device is not intended for patient transport outside a healthcare facility.

* The device is not intended for public use.

* This device is intended for no-invasive measuring and monitoring of arterial blood pressure.It is not 

intended for use on extremities other than the arm or for functions other than obtaining a blood pressure 

measurement.

* Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood 

pressure.Do not begin or end medical treatment without asking a physician for treatment advice.

* If you are taking medication,consult your physician to determine the most appropriate time to measure 

your blood pressure. Never change a prescribed medication without consulting your physician.

The signs below might be in the user manual, labeling or other component. 

They are the requirement of standard and using.

Read the instructions (actual
symbol colours are white on a

blue background).

Symbol for “MANUFACTURER”

Symbol for “SERIAL NUMBER”

Symbol for “TYPE BF APPLIED 
PARTS”

Symbol for “DIRECT CURRENT”

Symbol for “ENVIRONMENT 
PROTECTION - Electrical waste 
products should not be disposed of 
with household waste. Please recycle 
where facilities exist. Check with your 
local authority or retailer for recycling 
advice”

Symbol for “Authorised Representative 
in the European Community”

EC REP

Symbol for “MANUFACTURE 
DATE”

Classification for water ingress
and particulate matter

Caution: These notes must be 
observed to prevent any damage 
to the device.

SN

0197

This product complies with 
MDD93/42/EEC requirements. 

Safety Information

Содержание WBP1

Страница 1: ...Blood Pressure Monitor Instruction Manual WBP1...

Страница 2: ...MENT 15 16 Check the memory Delete the Readings INFORMATION FOR USER 17 Tips for measurement Maintenance ABOUT BLOOD PRESSURE 18 19 What are systolic pressure and diastolic pressure What is the standa...

Страница 3: ...in measuring blood pressure and heartbeat rate with arm circumference 22cm to 42cm about 8 16 It is intended for adult indoor use only 1 The device is not suitable for use on may be pregnant women or...

Страница 4: ...st be treated with caution Observe the storage and operating conditions described in this booklet The equipment is not AP APG equipment and not suitable for use in the presence of a flammable anesthet...

Страница 5: ...d sales service centers Please report to Harvard Medical Devices if any unexpected operation or events occur Keep the unit out of reach of infants young children or pets to avoid inhalation or swallow...

Страница 6: ...is dim The display does not light up 1 Battery powered mode 6VDC 4 AAA batteries Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batterie...

Страница 7: ...is stored in the memory The setting range of the year 2017 2057 Time format 24H the YEAR Each press will increase the numeral by one in a cycling manner mmHg mmHg min for 3 seconds to enter the mode...

Страница 8: ...same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first measuring Wait at least 3 minutes between measurements This allows you...

Страница 9: ...off otherwise it will turn off within 1 minute 1 When the monitor is off please press button to show the average value of the latest three records If the records are less than 3 groups it will displa...

Страница 10: ...ken in the following circumstances 1 Within 1 hour after drinking or eating 2 Immediate measurement after tea coffee or smoking 3 Within 20 minutes after taking a bath 4 When talking or moving your ng...

Страница 11: ...records all the pulse intervals and calculate the average if there are two or more pulse intervals the difference between each interval and the average is more than the average value of 25 or there a...

Страница 12: ...nd frequently asked questions for problems you may encounter with your blood pressure monitor If the product is not operating as you think it should check here before arranging for servicing Type BF a...

Страница 13: ...cation of usability engineering to medical devices EN 62304 2006 AC 2008 IEC 62304 2006 A1 2015 Medical device software Software life cycle processes Risk management Labeling User manual General Requi...

Страница 14: ...lt Technical description Table 1 Group I Class B Not application Not application 1 all necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFOR MANCE with regard to electromagnetic dis...

Страница 15: ...2 0 3 9 810 870 930 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse m...

Отзывы: