14. EMC Table
Guidance and manufacture’s declaration – electromagnetic emission
Emission test
RF emission
Harmonic
Voltage
Group 1
Class B
Not applicable
Not applicable
The Compressor Nebuliser use RF energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
The Compressor Nebuliser is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
CISPR 11
RF emissions
emissions
fluctuations/
flicker emissions
IEC 61000-3-3
IEC 61000-3-2
CISPR 11
Compliance
Electromagnetic environment – guidance
The Compressor Nebuliser is intended for use in the electromagnetic environment
specified below. The customer or the user of the Compressor Nebuliser should
assure that it is used in such an environment.
Guidance and manufacture’s declaration
Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
Power frequency
magnetic field IEC
61000-4-8
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
3V/m
3V/m
Floors should be wood, concrete or
ceramic tile. If floor are covered with
synthetic material, the relative humidity
should be at least 30%.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
(50Hz/60Hz)
Electrostatic
IEC 61000-4-2
discharge (ESD)
IEC 60601 test
Compliance
level
level
Electromagnetic environment - guidance
The Compressor Nebuliser is intended for use in the electromagnetic environment specified
below. The customer or the user of Compressor Nebuliser should assure that it is used
in such an environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.