DISPLAY THE MINIMUM AND THE MAXIMUM

Once the measurement is frozen :

Press on «

 Hold/min/max 

». 

The device displays the maximum value of pressure measured during the last questioning at the top of the screen and the minimum value of 
pressure measured since the last questioning at the bottom of the screen. 

Press again on « 

Hold/min/max 

».

The device displays the maximum value of velocity measured since the last questioning at the top of the screen and the minimum value of 
velocity since the last questioning at the bottom of the screen.

Press on

 « 

On/Off/Esc

 » to returns to the measurement display

.

ADJUST THE DEVICE

Select the measuring unit in pressure :

The device is on and displays the measurement.

Press on « 

Select

 » until « 

UNIT 

» appears on screen with the pressure unit indicated at the bottom of the screen.

Press on « 

OK 

».

The unit currently used blinks on screen.

Press on « 

Select

 » until the required unit appears : Pa, mmh2O, inWg or daPa

.

Press on « 

OK

 » to validate the unit selection.

« 

UNIT 

» blinks on screen.

Press on « 

On/Off/Esc

 » to return to the measurement display. 

Device drop-down menu 

 SEL.

          

BEEP

        

        OFF

            

 SEL.

          

COEF

    1.0015

                

 

            

Select

8x

 SEL.

       

 P_ATM

         1013

                

 hPa

            

 SEL.

           

UNIT

        

           Pa

            

3x

4x

 SEL.

          

VELO

        

          ON

            

1x

Select

2x

5x

6x

7x

 SEL.

           

UNIT

        

       Km/h

            

Select

Select

Select

Select

Select

 SEL.

         

AUTO

       OFF

            OFF

Select

 SEL.

          

TEMP

          20.0

                

  °C

            

OK

OK

Select

 SEL.

           

UNIT

        

           Pa

            

 SEL.

           

UNIT

        

       daPa

            

            0.1

daPa

          0.40

           Km/h

On/Off/

Esc

           

UNIT

SEL.

                  Pa

            

           

UNIT

SEL.

    Pa / daPa...

            

Measure and display the velocity :

It's possible to measure and display the velocity on a MP120 manometer. A differential pressure element (pitot tube, debimo blades,...) 
must be necessarily connected to the manometer.

The device is on and displays the pressure measurement.

Press on « 

Select

 ».

« 

VELO 

» blinks on screen.

Press on « 

OK

 ».

« 

OFF 

» blinks on screen.

Press on « 

Select

 » to select « 

ON 

» then press on « 

OK

 » to validate

.

« 

VELO

 » blinks on screen.

Press on « 

On/Off/Esc

 » to return to the measurements display.

OK

OK

Select

 SEL.

          

VELO

        

          ON

            

 SEL.

          

VELO

        

        OFF

            

            0.1

daPa

          

On/Off/

Esc

         

VELO

SEL.

                 ON

            

         

VELO

SEL.

       ON / OFF

            