Care and maintenance
To keep the balance operating properly, the housing and platform should be kept clean and free from foreign
material. If necessary, a cloth dampened with a mild detergent may be used. Keep calibration masses in a safe dry
place.
Unplug the AC Adapter when not in use. For long term storage, remove the batteries.
Troubleshooting
SYMPTOM
PROBABLE CAUSE
REMEDY
No Display
Power Adapter not connected.
d d
Connect power Adapter.
l
b
LO BAT Indicator
Batteries are weak
Replace batteries
Incorrect Weight Reading
Balance was not re-zeroed before
weighing Balance has not been
Press ON TARE with no weight on
the platform, then weigh item.
Calibration Procedure does not
work
Incorrect weights being
Used
Use correct weights.
Unable to display Weight in a
particular weighing unit
Weighing unit not
Enabled in Setup menu.
Use Unit Selection submenu to set
desired units ON (see Setup menu).
Unable to access C.
UNIT in Setup menu
to enter Conversion Factor
Custom Units (%) not
Enabled in Setup menu, or not set to
USR.
Use Unit Selection submenu to set
Custom
Units (%) to USR.
Balance won’t store
Selections made in Setup menu
END selection was not
used to exit Setup menu.
You MUST use END to
Exit Setup menu and
Save selections.
Error codes
The following list describes the various errors which can appear on the display and the suggested remedy.
•
Err 0: With the platform on, this indicates the platform is too light, make it heavier properly. If this error
still occurs, the sensor or the circuit board may be destroyed. The balance must be sent to the
manufacturer to be serviced.
•
Err 1: Sample being weighed exceeds the capacity of the balance. If error occurs when the sample is within
the balance capacity, balance may be incorrectly calibrated.
•
Err 2: This indicates there is no platform or the platform is too light when turning the balance on. Please
put the platform on before turning on the balance.