21
4
、
Setting and measuring mode for KTM-150/KTM-15
4.1 Mode setting
Press the MODE key over one second.
It means getting into setting mode when the words PP----PD shown on the below left on LCD screen.
Please press the MODE key if needing to change measuring mode:
maximum value (PP)----only indicating the first peak value (PD)----real-time value (C)----for calibration (TRACK).
Measurement mode
Display
Contents
P-P
(peak to peak)
Display section
「
PP
」
The maximum under measurement is always displayed.
Load holds more than 10 digits*.
Usually, this mode is used.
P-D
(peak-down)
Display section
「
PD
」
A value when a load torque value changes from a rise to
descent is displayed. It carries out, once it pushes a clear
button, when applying re-load.
Load operates more than 10 digits*.
It is suitable for measurement of a torque wrench etc.
C
(Real-time Output)
Display section
「
C
」
The data output of the load torque value is carried out
every about 1 / 180 seconds.
(A change whole 1 / 12 seconds is possible)
It is suitable for torque curve creation or a screw bundle
examination.
TRACK
There is no display
It is mainly used at the time of calibration etc.
The value of the load torque concerning a sensor is
displayed as it is.
(
There is no display.)
* digit–10 digits is 0.10 N
.
m. (1.0 kgf
.
cm/1.0 lbf
.
in)
4.2 Test unit selection
Press UNIT key over one second, the system will get into setting mode. The unit will change whenever user press UNIT key.
lbf.in-------------kgf.cm-------------N.m-------------(cycle)
4.3 Power capacity indication
Displaying on LCD
Contents
Sufficient power capacity, no need to recharge, the power capacity is around 100% left.
The power capacity is around 50% left.
It is not allowed to use, needs to urgent charge when the power capacity is 10% left.
Because the power will be shut off anytime.
The power cannot be on when the power capacity is 0% left.
LCD will be not working also.
Please use original charger and turn off the power when charging. (input: 100V~ 240V; output: 12V 500mA)
5
、
The operation of functional setting key
:
The operation of functional setting key should collocate with UP key and DOWN key to operate the function except its own function.
Setting proper value range of torque
First of all, setting the torque value both minimum limit and maximum limit, so that user can judge the difference of torque value.
For example: We need to test 5 kgf.cm torque. We can set a tolerated range 0.5 kgf.cm, so we can set minimum limit with 4.5 kgf.cm;
maximum limit with 5.5 kgf.cm.
If the testing value is less than 4.5 kgf.cm or higher than 5.5 kgf.cm, the NG light will be on. On the other hand, the GOOD light will
be on if the testing value is within the range.
PRG/SET
○
UP
○
A numerical value is increased.
DOWN
○
A place is changed
.
A numerical value
is determined.
Содержание KTM-15