48 Specifications
TOS8030
9.2
Other Functions
9.3
General Specifications
Remote control
Connector
5-pin DIN connector on the rear panel
Control available
Start/stop of testing
Optional devices connectable
Remote control boxes: RC01-TOS and RC02-TOS
High-voltage test probes: HP01A-TOS and HP02A-TOS
Signal I/O
Connector
14-pin screw-less terminal on the rear panel
Status signal output
Output of a READY signal / H.V ON signal / PASS signal / FAIL signal
/ PROTECTION signal
Make-contact signal (contact rating: 1 A/30 Vdc)
INTERLOCK input terminal
A PROTECTION state is invoked when the terminals are open.
Environment
Operation environment
Indoor use, Overvoltage Category II
Specifications-
assured range
Temperature
5
°
C to 35
°
C
Relative humidity 20% to 80% (with no dew condensation)
Operating range
Temperature
0
°
C to 40
°
C
Relative humidity 20% to 80% (with no dew condensation)
Storage range
Temperature
-40
°
C to 70
°
C
Relative humidity 90% or less (with no dew condensation)
Altitude
Up to 2000 m
AC input
Nominal input rating
(Input voltage range)
220 V (200 V to 240 V) or 100 V (90 V to 110 V),
50 Hz or 60 Hz
Power consumption
At no-load
(in READY state)
10 VA or less
At rated load
50 VA maximum
Insulation resistance
AC INPUT to
chassis
30 M
Ω
or more (at 500 Vdc)
Withstand voltage
AC INPUT to
chassis
10 mA or less when 1390 Vac is applied for 2 seconds
Earth continuity
25 Aac/0.1
Ω
or less
Dimensions (largest section)
160 (170) W x 132 (155) H x 230 (270) D mm
Weight
Approx. 6 kg
Accessories
•
High-voltage test leads TL01-TOS (approx. 1.5 m): 1 set
•
Power cord: 1
•
INTERLOCK jumper: 1
•
Operation Manual: 1 copy