TOS7200
25
Setting Limits
Setting the upper resistance (UPPER)
The upper resistance can be set in the range of 0.01 MΩ to 5000 MΩ (provided that the current
is within the maximum rated current).
1
Press the UPPER/LOWER key.
This causes the rightmost digit of the resistance meter LED to blink, indicating
that the cursor is positioned at this digit.
2
Check whether the UPPER LED is lit.
If it is, the upper resistance is displayed. If the LOWER LED is lit, the value
displayed is the lower resistance. In such a case, press the UPPER/LOWER key
again to light up the UPPER LED.
3
Press the or key to move the cursor to the digit at which a value is to
be set.
4
With the SHIFT key held down, press the or key to change the value.
If necessary, change the value at another digit to reach the desired upper
resistance.
Upper Resistance
Resolution
0.01 MΩ to 9.99 MΩ
0.01 MΩ
10.0 MΩ to 99.9 MΩ
0.1 MΩ
100 MΩ to 5000 MΩ
1 MΩ
When both the upper and lower judgments are activated, setting the upper resistance to a
value below the lower resistance causes the UPPER LED to blink, notifying the operator that
the test cannot be conducted. (The upper resistance is set to 100 MΩ at factory shipment.)
In such cases, raise the upper resistance, lower the lower resistance, or set the lower
judgment to OFF.