TOS5200
47
CONFIG Settings
5
W
ithst
a
nding
V
o
lt
age
T
e
st
Test
Sets whether the 24 Vdc output is generated while tests are being performed with the test
voltage at the specified voltage value.
The 24 Vdc output is not generated during voltage rise time (Rise Time) and voltage fall time
(Fall Time).
Pass
Sets whether the 24 Vdc output is generated after a PASS judgment has occurred.
Upper Fail
Sets whether the 24 Vdc output is generated after an upper limit FAIL judgment has occurred.
Lower Fail
Sets whether the 24 Vdc output is generated after a lower limit FAIL judgment has occurred.
Ready
Sets whether the 24 Vdc output is generated while the READY LED is lit.
Protection
Sets whether the 24 Vdc output is generated while the TOS5200 is in protection mode.
Power ON
Sets whether the 24 Vdc output is generated while the POWER switch is on.
off (0)
The 24 Vdc output is turned off during testing (factory default setting).
on (1)
The 24 Vdc output is turned on during testing.
off (0)
The 24 Vdc output is turned off during testing (factory default setting).
on (1)
The 24 Vdc output is turned on during testing.
off (0)
The 24 Vdc output is turned off after an upper limit FAIL judgment occurs
(factory default setting).
on (1)
The 24 Vdc output is turned on after an upper limit FAIL judgment occurs.
off (0)
The 24 Vdc output is turned off after an lower limit FAIL judgment occurs
(factory default setting).
on (1)
The 24 Vdc output is turned on after a lower limit FAIL judgment occurs.
off (0)
The 24 Vdc output is turned off while the READY LED is lit (factory default
setting).
on (1)
The 24 Vdc output is turned on while the READY LED is lit.
off (0)
The 24 Vdc output is turned off when the TOS5200 is in protection mode
(factory default setting).
on (1)
The 24 Vdc output is turned on when the TOS5200 is in protection mode.
off (0)
The 24 Vdc output is turned off when the POWER switch is turned on (fac-
tory default setting).
on (1)
The 24 Vdc output is turned on when the POWER switch is turned on.
Содержание TOS5200
Страница 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Страница 18: ...18 TOS5200 This page is intentionally blank ...
Страница 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Страница 31: ...Panel Operation Basics This chapter describes how to select items on the screen and how to enter values ...
Страница 60: ...60 TOS5200 This page is intentionally blank ...
Страница 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Страница 63: ...Specifications This chapter contains the specifications and gives the dimensions of the TOS5200 ...
Страница 71: ...Appendix A List of Default Settings B Protection Functions C Protection Functions D Troubleshooting ...
Страница 78: ...78 TOS5200 This page is intentionally blank ...
Страница 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Страница 81: ......