24
TOS5200
Connecting to the Device under Test (DUT)
Using test leads
1
Check that the POWER switch is off and that the DANGER LED is off.
2
Check that there are no tears or breaks in the test lead insulation.
3
Raise the front-panel LOW VOLTAGE terminal’s cable lock, and then
connect the low-voltage test lead (black).
4
Lower the cable lock to secure the lead in place.
WARNING
Risk of electric shock. During testing (while the TEST LED or DANGER LED is lit),
never touch the HIGH VOLTAGE terminal, test leads, and DUT.
WARNING
Risk of electric shock.
• Parts of the included test leads near the alligator clips
protrude from the vinyl insulation when the wires are
connected. These parts are dangerous. Never come
close to these parts during testing.
• If connections are incomplete, the entire DUT may be
charged to a high voltage. This is dangerous, so be sure
to connect the DUT correctly.
• Be sure to connect the low-voltage test lead (black) first.
Never come close to this area
during testing.
OUTPUT
DANGER
TOS5200
Off
Off
POWER
See
HIGH VOLTAGE
Low-voltage test lead (black)
Raise the cable lock.
Low-voltage test lead (black)
Lower the cable lock.
Содержание TOS5200
Страница 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Страница 18: ...18 TOS5200 This page is intentionally blank ...
Страница 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Страница 31: ...Panel Operation Basics This chapter describes how to select items on the screen and how to enter values ...
Страница 60: ...60 TOS5200 This page is intentionally blank ...
Страница 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Страница 63: ...Specifications This chapter contains the specifications and gives the dimensions of the TOS5200 ...
Страница 71: ...Appendix A List of Default Settings B Protection Functions C Protection Functions D Troubleshooting ...
Страница 78: ...78 TOS5200 This page is intentionally blank ...
Страница 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Страница 81: ......