92
PLZ-4WL
Controlling the Current Range Externally
You can use an external control signal to control the current range. You can use the range
status output to monitor the present state of the range.
You cannot control the voltage range.
Set the current range on the panel to the H range.
The current range cannot be changed when the load is turned on. The control signal input
received while the load is turned on is ignored.
Input for switching the range externally
Use pins 10 (RANGE CONT 1) and 11 (RANGE
CONT 0) of the J1 connector (use pin 15 or 16
as the common).
The signal is 2 bit.
The control terminals are pulled up to the
internal circuit’s +5 V level by a resistor that is
approximately 10 k
. The maximum allowable
voltage is 5 V. The logic threshold levels are
CMOS.
Range status output
Use pins 18 (RANGE STATUS 1) and
19 (RANGE STATUS 0) of the J1
connector (pin 22 is common).
The signal is 2 bit.
Current Range
Control Input
1
Status Output
2
PLZ
164WL
PLZ
334WL
RANGE
CONT 0
RANGE
CONT 1
RANGE
STATUS 0
RANGE
STATUS 1
H
50 A
100 A
HIGH
HIGH
OFF
OFF
M
5 A
10 A
HIGH
LOW
OFF
ON
L
0.5 A
1 A
LOW
HIGH
ON
OFF
1
HIGH: 5 V, LOW: 0 V
2
OFF: OPEN, ON: SHORT
+5 V
Approx. 10 kΩ
A COM
J1
15,16
10,11
J1
Photocoupler (P.C)
22
18 , 19
Maximum applied voltage: 30 V
Maximum current:
8 mA
Содержание PLZ164WL
Страница 11: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 25: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 43: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 57: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 79: ...External Control This chapter explains external control 5 ...
Страница 95: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 6 ...
Страница 107: ...7 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WL ...
Страница 117: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...