PLZ-4WL
35
2
Basi
c Funct
ions
Constant Power Mode (CP mode)
In constant power mode (CP mode), you set the power [W].
Configuring CP mode
Select the operation mode, and set the power.
1
Make sure that the load is off.
The load is off when the LED to the upper left of the LOAD key is off. If the LED is lit,
press LOAD to turn the load off.
2
Press MODE to select the operation mode (CP).
The operation mode popup menu appears.
Press MODE repetitively until CP is highlighted on the menu. After you select the
operation mode, the popup menu disappears, and “CP” appears on the display.
3
Press RANGE to select the current range.
The current range popup menu appears.
Each time you press RANGE, the range switches between L, M, and H. Press RANGE
until the range that you want to select is highlighted. After you select a range, the
popup menu disappears, and the range that you selected appears along with its full
scale value. The full scale value varies depending on the model.
4
Press VRANGE (SHIFT+RANGE) to select the voltage range.
The voltage range popup menu appears.
Each time you press VRANGE (SHIFT+RANGE), the voltage range switches between
4 V and 30 V. Press VRANGE (SHIFT+RANGE) until the range that you want to select
is highlighted. After you select a voltage range, it appears on the display.
5
Check that the display is in the fundamental setting entry condition.
If SET is not highlighted, press SET/VSET to switch to the fundamental setting entry
condition.
6
Turn the rotary knob to set the power.
The CP mode settings are complete. When you press LOAD, the LOAD LED lights, and the
PLZ-4WL starts supplying current. You can change the voltage even when the load is on.
S
W
ITCHING
ENTER
SHORT
EDIT
SLE
W
RATE
RUN / STOP
UVP
PAUSE
STORE
RECALL
A
B
C
LOAD
Th/TL
FREQ / DUTY
MENU
PRESET SEQ
I
LEVEL
VRANGE
RANGE
S
W
ON
+CV
MODE
LOAD
SET / VSET
LOCK
SHIFT
ABC
OPP / OCP
Memo
When “SET” is
highlighted, the PLZ-4WL
is in the fundamental
setting entry condition.
Содержание PLZ164WL
Страница 11: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 25: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 43: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 57: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 79: ...External Control This chapter explains external control 5 ...
Страница 95: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 6 ...
Страница 107: ...7 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WL ...
Страница 117: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...
Страница 137: ......