PLZ-4WH
Short
When the short function is enabled, the maximum current is set in CC mode or the minimum
resistance is set in CR mode, and the short signal output of the EXT CONT connector is
turned on. The short signal output terminal is a relay contact (30 Vdc/1 A). You can drive an
external relay for high currents or a similar device and short the load input terminals.
Turning the short function on and off
You can turn the short function on and off when the load is on.
.
If you press SHORT (SHIFT+SLEW RATE) while the short function is off, the short icon is
displayed, and the short function turns on. In constant current mode (CC mode), the
fundamental setting is set to the maximum current value; in constant resistance mode (CR
mode), the fundamental setting is set to the minimum resistance value.
If you press SHORT (SHIFT+SLEW RATE) while the short function is on, the short icon
disappears, and the short function turns off. The fundamental setting returns to the value that
it was at before the short function was turned on.
See
Be sure to use a dedicated driver circuit to drive a relay for high current. Please supply the
dedicated driver circuit yourself.
+
DUT
PLZ-4WH Series
Relay
High-current relay
EXT CONT connector
on the rear panel
-
+
-
<19>
<20>
EXT CONT
Driver circuit
SWITCHING
ENTER
SHORT
EDIT
SLEW RATE
RUN / STOP
UVP
PAUSE
STORE
RECALL
A
B
C
LOAD
Th/TL
FREQ / DUTY
MENU
PRESET SEQ
I
LEVEL
VRANGE
RANGE
SW ON
+CV
MODE
LOAD
SET / VSET
LOCK
SHIFT
ABC
OPP / OCP
Short icon
0.0000
A
0.00
V
0.00
W
CC
SLEWRATE
0.5000
A
132.00
mA/µs
SET
H 8.25A 650V
Содержание PLZ1004WH
Страница 13: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 28: ...28 PLZ 4WH Remote Sensing ...
Страница 29: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 47: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 61: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 83: ...External Control and Parallel Operation This chapter explains external control and Parallel Operation 5 ...
Страница 115: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 7 ...
Страница 127: ...8 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WH ...
Страница 137: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...