PLZ-4W
Installation and Preparation 2-15
Connection procedure of the load input terminal on the rear panel
• Do not touch the load input terminal while the PLZ-4W is turned ON, as it
may lead to electric shock.
In addition, be sure to use the load input terminal cover.
• The load input terminals on the front and rear panels are coupled inside
the PLZ-4W. A voltage applied to one end will appear on the other end.
• Do not connect equipment to the load input terminals on the front and rear
panels simultaneously, as it can cause damage.
• There is a danger of breakdown. Do not connect the equipment under test
to the load input terminal when the load is turned on.
• There is a danger of overheat. Attach crimping terminal to the wire and
use the set of screws that came with the package for connection.
1.
Turn off the POWER switch.
2.
Check that the output of the equipment under test is off.
3.
Connect the load wire to the load input terminal on the rear panel.
Use the load input terminal cover. See Fig. 2-13 on how to use the cover.
For the connection procedure of the load wire, see Fig. 2-15.
4.
Fix the load input terminal cover to the rear panel using the lock plate.
Pass the inner pin of the lock plate through the hole on the side of the cover.
For details, see Fig. 2-16.
5.
Connect the load wire to the output terminal of the equipment under
test.
6.
Check the polarity of the connection.
Fig. 2-15
Connection to the load input terminal on the rear panel
WARNING
CAUTION
Bolt (M8
×
18)
Spring washer (M8)
Load input terminal cover
Nut (M8)
Crimp terminal
Be sure to use the
set of screws
Содержание PLZ-4W Series
Страница 10: ...VIII Contents PLZ 4W ...
Страница 20: ...1 10 General Information PLZ 4W ...
Страница 38: ...2 18 Installation and Preparation PLZ 4W ...
Страница 90: ...5 30 Basic Operation PLZ 4W ...
Страница 192: ...9 10 Specifications PLZ 4W ...
Страница 214: ...A 22 Appendix PLZ 4W ...