2-10 Installation and Preparation
PLZ-4W
2.8
Load Wiring
To ensure that the functions of the PLZ-4W work accurately and reliably, all wires
must be connected correctly to their loads.
• This operation manual refers to the terminal on the rear panel to which the equip-
ment under test is connected and current is supplied as load input terminal.
2.8.1
Precautions Concerning Wiring
Electric wire used
• Use a load wire with suf
fi
cient diameter for the current as well as non-
fl
am-
mable or
fl
ame-resistant cover.
If the resistance of the load wire is large, a large voltage drop may occur when a cur-
rent is supplied and the voltage at the load input terminal may fall below the mini-
mum operating voltage. Refer to Table 2-1 and select thick wires as much as possible.
Table2-1
Nominal cross-sectional area of wires and allowable currents
Nominal Cross-
Sectional Area
[mm
2
]
AWG
(Reference Cross-
Sectional Area)
[mm
2
]
Allowable Current(*)
[A]
(Ta = 30
°
C)
Kikusui-Recom-
mended Current
[A]
2
14
(2.08)
27
10
3.5
12
(3.31)
37
-
5.5
10
(5.26)
49
20
8
8
(8.37)
61
30
14
6
(13.3)
88
50
22
4
(21.15)
115
80
30
2
(33.62)
139
-
38
1
(42.41)
162
100
50
1/0
(53.49)
190
-
60
2/0
(67.43)
217
-
80
3/0
(85.01)
257
200
100
4/0
(107.2)
298
-
125
-
-
344
-
150
-
-
395
300
200
-
-
469
-
250
-
-
556
-
325
-
-
650
-
*Excerpts from Japanese laws related to electrical equipment.
NOTE
CAUTION