Sequence Example (Normal Sequence)
Programs are created using the PLZ-4W operation panel. This section describes the
procedure of entering the following example sequence from the operation panel.
• Flow of the example sequence
In this example, we will assume a sequence that executes 2 programs (program
numbers 1 and 2) to simulate the waveform of Fig. 6-15.
As shown in Fig. 6-15, program 1 can be divided into 3 steps. After the execution of
the third step of program 1, the sequence continues to program 2. Program 2 exe-
cutes the 3 steps of Fig. 6-15 and ends the
fi
rst loop. Here, we will set program 2 to
be executed twice. When program 2 is executed 2 times, this sequence operation is
fi
nished.
Fig. 6-15
Sequence execution example
Program 1
CC mode (executed once)
1) 7 A
200 s
LOAD ON
2) 7 A
150 s
LOAD ON
3) 0.5 A
80 s
LOAD OFF
Program 2
CC mode (executed twice)
1) 10 A
200 s
LOAD ON
2) 5 A
50 s
LOAD ON
3) 8 A
150 s
LOAD ON
TRIG
TRIG OFF
TRIG OFF
TRIG
ON
TRIG OFF
TRIG OFF
TRIG
OFF
TRIG OFF
TRIG OFF
TRIG
OFF
10A
7A
8A
5A
Executio (RUN)
7A
200 s
LOAD ON
7A
150 s
LOAD ON
0.5A
80 s
LOAD OFF
10A
200 s
LOAD ON
5A
50 s
LOAD ON
8A
150 s
LOAD ON
10A
200 s
LOAD ON
5A
50 s
LOAD ON
8A
150 s
LOAD ON
Step 1
Program 1
Program 2
Program 2
Step 1
Step 2
Step 2
Step 3
Step 3
Step 1
Step 2
Step 3
Input current
CC