RC03-PCR-LA
Operation Method 4-3
4.3
Functions Extended by Using the
Remote Controller
Some functions are added to the PCR-LA AC power supply by using
the Remote Controller. This section describes how to use such func-
tions.
Power Line Abnormality Simulation
This function allows the PCR-LA power supply to simulate a inter-
ruption, fast voltage swell (pops), or fast voltage dip (dips) to conduct
a power line abnormality simulation in the AC mode.
Fig.4-2
Simulated Power Line Abnormality Wave-
forms
A power line abnormality simulation should be conducted after
parameters are set. Press the SIM key, select the desired item from
among T1 to T3 and others shown in Fig. 4-3, and then set a value
using the numeric keys or rotary encoder (for setting using the
numeric keys, press the ENTER key to con
fi
rm the value). Press the
RUN key to execute the simulation. To stop the simulation, press the
STOP key. The V (T3) key sets an rms value.
Fig.4-3
Simulated Waveforms and Parameters
Power interruption
waveform
Fast voltage swell
(pop)
Fast voltage dip
(dip)