Fine adjustment function
This function increases the resolution of the VOLTAGE and CURRENT knobs. By
turning the VOLTAGE or CURRENT knob while holding down the SHIFT key,
you can set the value using finer resolution.
The display resolution of the preset voltage and preset current does not change even
if you use the fine adjustment function. Therefore, the displayed value may not
change even if you turn the knob. The display changes when the amount of change
reaches the minimum displayed digit of the preset voltage or current.
4.6
Protection Functions and Alarms
The PAT is equipped with the following protection function.
•
Overvoltage protection (OVP)
•
Overcurrent protection (OCP)
•
Overheat protection (OHP)
•
Overheat protection of the bleeder circuit (BOHP)
•
Fan failure protection (FAN)
•
Incorrect sensing connection protection (SENSE)
•
Input open-phase protection (PHASE)
•
Shut down (SD)
Alarm occurrence and release
Alarm occurrence
When a protection function activates, the PAT behaves as follows:
•
The output turns off.
For the overvoltage protection (OVP), overcurrent protection (OCP), and
shutdown (SD), you can select breaker trip in the CONFIG settings.
to p. 4-14
•
The ALARM LED on the front panel display illuminates, then the cause of alarm
occurrence and the current measuring value displays alternately.
When the breaker trip is selected in overvoltage protection (OVP) or overcurrent
protection (OCP) or shutdown (SD), the front panel display shows the characters
indicated as power off respectively for about 10 to 15 seconds.
•
The ALARM signal is output from pin 5 of the J1 connector.
The ALARM signal is delivered 10 to 15 seconds even if the breaker is tripped.
See
See
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...