8.3
Calibration
The PAT is shipped after carrying out appropriate calibrations. We recommend periodic
calibration to maintain the performance over an extended period.
For calibration, contact your Kikusui agent or distributor.
If you are going to calibrate the PAT yourself, follow the procedures below.
The calibration procedures given in this section includes all calibration items of the PAT.
8.3.1
Calibration Overview
The following six calibration items are available.
•
Output voltage and voltmeter
•
Output current and ammeter
•
Overvoltage protection
•
Overcurrent protection
•
Output voltage control
•
Output current control
Test Equipment Required
For calibration, the following equipment is necessary.
•
DC voltmeter (DVM) with measuring accuracy of 0.02 % or better.
•
Shunt (See Table 8-1)
Table
8
-1
Recommended sh
u
nt resistors
•
10-k
Ω
resistor with accuracy of 1 % or better
•
DC power supply (variable power supply that can output +10 V, used as a
voltage source)
Environment
Perform calibration under the following environment.
•
Temperature: 23 ° C ± 5 ° C
•
Relative humidity: 80 % or less
To minimize the calibration error due to initial drift, warm up the PAT for at least 30
minutes before calibration. In addition, warm up the DVM and shunt resistor
adequately.
Model
Sh
u
nt resistors
Rating
Tolerance
PAT20-200T
200 A/50 mV (0.25 m
Ω
)
<
±
0.1 %
PAT40-100T
100 A/50 mV (0.5 m
Ω
)
PAT60-67T
PAT160-25T
50 A/50 mV (1 m
Ω
)
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...