3-12 Operation
KJM6310
Adjustment procedure
1. Connect a DC voltmeter to the JITTER OUT connector.
Send a DELAY MODE message and set the MANUAL delay
through the RS-232C interface.
2. Send a DELAY CONTROL message through the RS-232C
interface.
3. Vary the value of the command so that the DC voltmeter
reading is as close to 0 as possible.
The phase difference is at the optimum value when the voltmeter
reading is 0 V.