6-22
KHA1000
6.4
Report Printout Format
6.4.1
Printing Reports
The following figures show the formats for printing reports. Fig.6-7 shows an
example of d measurement and concurrent measurement of Pst and Plt (flicker).
Fig.6-8
shows an example of the method complying with the “Test conditions and
procedure for measuring d max voltage changes caused by manual switching.”
Items differ depending on whether the nominal voltage is 230 V or 100 V.
Each item in the figure is explained in Table 6-2. The numbers in the table
correspond to those in Fig.6-7 and Fig.6-8.
Fig.6-7
Example of report printout (d measurement and
concurrent measurement of Pst and Plt (flicker))
See
Company
Model name
PXZ1600
Type
Prototype
Serial No.
12345678
Operating mode
Date of test
2021/4/16 18:25:18
Climatic condition
Memo
Example
Power analyzer
Supply Source
Reference Impedance
Test Data of Voltage Fluction and Flicker
FINAL TEST RESULT
PASS *1*2
Nominal Voltage
100 V
Nominal Frequency
50 Hz
Plt Test duration
150 s
Voltage Fluctuation and Flicker TEST REPORT
*1 Limits apply for the result at the condition of nominal voltage 230V and nominal frequency 50Hz.
Test Engineer
Type of test
Flicker Margin
100 %
d Measurement Margin 10 %
Pst
dc(%)
dmax(%)
Tmax(ms)
Judge
Limit
Plt
Judge
Limit
Measurement
1.000
3.300
4.000
500
Seg. 1
0.207
0.210
0.500
0
Warning
Seg. 2
0.206
0.130
0.560
0
Warning
Seg. 3
0.206
0.180
0.630
0
Warning
Seg. 4
0.219
0.190
0.540
0
Warning
Seg. 5
0.163
0.170
0.380
0
PASS
Seg. 6
-.---
-.---
-.---
------
Seg. 7
-.---
-.---
-.---
------
Seg. 8
-.---
-.---
-.---
------
Seg. 9
-.---
-.---
-.---
------
Seg.10
-.---
-.---
-.---
------
Seg.11
-.---
-.---
-.---
------
Seg.12
-.---
-.---
-.---
------
0.650
0.151
PASS
*2 The values for the final test result are more than Margin value and less than or equal to Limit value.
EN 61000-3-3(2013)/A1(2019)
KHA1000,Ver2.00
*IEC 61000-3-3:Ed3.1(2017)
EN 61000-4-15(2011)
*IEC 61000-4-15:Ed2.0(2010)
S
J
d
-
T
J
1
2
4
3
5
6
7
8
9
10
11
ASS *1*2
P
12
13
)
T
9)
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......