6-24
KHA1000
Table 6-2
Explanation of report items
Explanation
No.
Item
Content
Explanation
1
Company
Optional descriptions such as
company names (underlined)
Description column after forms output
Test Engineer
Optional descriptions such as test
engineers (underlined)
Model Name
Model name of EUT
Data input by view type of Vf-
Observation and Analysis display
results list, voltage fluctuation list
(manual)
view setting
Type
Type of EUT
Serial No.
Serial No. of EUT
2
Operating mode
Optional descriptions such as
operating mode of EUT (underlined)
Description column after forms output
Date of test
Test execution date/time
Data input when setting date/time in
system setting display
Climatic condition
Optional climatic descriptions such as
temperature (underlined)
Description column after forms output
Memo
Memo
Data input by view type of Vf-
Observation and Analysis display
results list, voltage fluctuation list
(manual)
view setting
3
Type of test
Test standard
Data input when setting the test
conditions
Power analyzer
Information about measuring
instrument used in tests
Model name and firmware version of
this product
Supply source
Optional descriptions such as
information about AC power supply for
tests. (Underlined).
Description column after forms output
Reference
Impedance
Information on reference impedance: fixed to (0.40 + j0.25) ohms
4
FINAL TEST
RESULT
Final judgment result: PASS or FAIL
Even in the case of PASS, *2 is displayed if the value exceeds the margin.
Nominal voltage
Nominal voltage
Nominal voltage input when setting
the test conditions
Nominal frequency Nominal frequency
Nominal frequency input when setting
the test conditions
Plt Test duration
Total measuring time
(= total test period)
Pst measurement time × Pst
measurement count input when
setting the test conditions
Measuring Time
d measurement time × d
measurement count input when
setting the test conditions
Flicker Margin
Flicker margin
Margin for the Pst and Plt limit values
input when setting the test conditions
d Measurement
Margin
d margin
Margin for the dc, dmax, and Tmax
*1
limit values input when setting the test
conditions
5
Limits apply for the
result...
*1 is displayed when the limit value at nominal voltage 230 V and nominal
frequency 50 Hz is applied.
The values for the
final test result are
more than Margin
value and less than
or equal to Limit
value.
*2 is displayed when the margin is exceeded even in the case of PASS.
6
Seg
Pst measurement time or d
measurement time corresponds to
one-segment (Seg1 -) time.
Pst measurement time or d
measurement time input when setting
the test conditions
7
Pst
Short-time flicker value in one-segment (Seg1 -) time
•
Limit: Limit value
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......