TROUBLESHOOTING
If your questions can’t be answered with below information, please contact your local agent directly. They might require a technician to take
care of the problem.
NO LIGHTS ON UNIT
• Check control unit is connect to the mains power supply and that the unit is
switched on.
• Check quick blow fuse on the control unit or mains plug fuse.
LOW PRESSURE ALERT
• Check main power to the control unit and that the plug is switched.
• Check mattress air supply connector is connected to the pump unit correctly.
Check CPR connector is connected properly.
• Under the mattress cover check all “T” & “L” connectors are connected to the
air tubes.
• Check the connector tubes for kinks, crimps or damage.
PATIENT’S BODY IS SAGGING IN THE
MIDDLE/ BOTTOMING OUT
Increase the value of the comfort the setting. Wait a couple of minutes so pres-
sure can stabilize before making another change.
CONTROL LOCK UP OR “FREEZE”
Turn the control unit off and unplug the pump. Wait a few minutes and plug the
control unit back in. Turn on the control unit.