Calibration Procedures
1
Keysight U3606B Service Guide
81
7
Press
to start the adjustment. The word CALib starts flashing to indicate
that the calibration is in progress.
– Successful completion of the adjustment is indicated by a short beep and
the display briefly showing PASS.
– An adjustment failure is indicated by a long beep, the display showing FAiL
along with a calibration error number. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the AC voltage gain adjustments using the
“AC voltage verification test”
Table 1-25
AC voltage gain adjustment
Adjustment item
Input voltage (V
rms
)
Input frequency
10.000 mV, 1 kHz
10 mV
1 kHz
100.000 mV, 1 kHz
100 mV
1 kHz
1.00000 V, 1 kHz
1 V
1 kHz
10.0000 V, 1 kHz
10 V
1 kHz
100.000 V, 1 kHz
100 V
1 kHz
750.00 V, 1 kHz
750 V
1 kHz
OUT
SBY
CAUTION
Set the calibrator output to 0 V before disconnecting from the instrument
input terminals.