1
Calibration Procedures
44
Keysight U3606B Service Guide
AC voltage verification test
1
Connect the calibrator to the front panel
V
(red) and
LO
(black) input terminals
(see
Figure 1-2, “Test setup for DC voltage and AC voltage gain verification,”
2
Press
[
]
to select the AC voltage function. The AC annunciator will turn on
according to your selection.
3
Select each range in the order shown below. Provide the indicated input
voltage and frequency. Compare the measurement results to the appropriate
test limits in
. (Be certain to allow for appropriate source settling
when using the Fluke 5520A.)
Table 1-5
AC voltage gain verification test
Input voltage
Input frequency
Range
Error from nominal 1-year
10 mV
1 kHz
100 mV
±0.12 mV
100 mV
1 kHz
100 mV
±0.3 mV
0.1 V
1 kHz
1 V
±0.0012 V
1 V
1 kHz
1 V
±0.003 V
1 V
1 kHz
10 V
±0.012 V
10 V
1 kHz
10 V
±0.03 V
10 V
1 kHz
100 V
±0.12 V
100 V
1 kHz
100 V
±0.3 V
100 V
1 kHz
750 V
±0.95 V
750 V
1 kHz
750 V
±2.25 V
V
CAUTION
Set the calibrator output to 0 V before disconnecting from the instrument
input terminals.
Содержание U3606B
Страница 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Страница 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 104: ...1 Calibration Procedures 104 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...