4
Performance Test
78
Keysight U3401A User’s and Service Guide
Resistance Verification Test
Configuration: 2-Wire
Ω
1
Select the resistance function.
2
Select each range in the order shown in
. Provide the resistance value
indicated. Compare measurement results to the appropriate test limits shown
in
. (Be certain to allow for appropriate source settling).
Table 4-4
2-wire
Ω
verification test
Function
Input
Range
Error from nominal one year
2-wire
Ω
0.00
Ω
500
Ω
±50 m
Ω
[a]
0.0000 k
Ω
5 k
Ω
±0.3
Ω
[a]
0.000 k
Ω
50 k
Ω
±3
Ω
0.00 k
Ω
500 k
Ω
±30
Ω
0.0000 M
Ω
5 M
Ω
±300
Ω
0.000 M
Ω
50 M
Ω
±3 k
Ω
450.00
Ω
500
Ω
±500 m
Ω
[a]
4.5000 k
Ω
5 k
Ω
±4.8
Ω
[a]
45.000 k
Ω
50 k
Ω
±48
Ω
450.00 k
Ω
500 k
Ω
±480
Ω
4.5000 M
Ω
5 M
Ω
±4.8 k
Ω
30.000 M
Ω
50 M
Ω
±93 k
Ω
Notes:
[a] Specifications are for 2-wire ohms function using the Rel operation enabled to eliminate lead resistance.
Содержание U3401A
Страница 1: ...Keysight U3401A 4 1 2 Digit Dual Display Multimeter User s and Service Guide ...
Страница 12: ...12 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...3 Application Tutorial 70 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 82: ...4 Performance Test 82 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...