U3042AE16
Cal Kit Operational Check
Verification Limits
Specifications for the test set are typical. System performance for the analyzer and test set are
only characteristic and intended as non-warranted information. Only a functional certificate is
provided for the test set.
It is recommended that you return your instrument to Keysight Technologies for servicing or
repair if the test set and network analyzer performance exceed the operational verification
limits.
A periodic calibration is not required. The Operators Check should be performed after System
Setup, or if performance is in question. An N-Port calibration should be performed before
making a measurement.
NOTE
Typical performance is based on 1 to 2 units.
Refer to
If you suspect that your 20-Port configuration is not operating properly, ensure
that all front RF jumper interconnect cables are correctly connected.
Table 17
PNA-L Reflection Tracking Limits
Frequency
Ports 1 - 4
Ports 5 - 20
300 kHz to 50 MHz
-9
-15
50 MHz to 500 MHz
-10
-16
500 MHz to 3.2 GHz
-12
-18
3.2 GHz to 10 GHz
-17
-23
10 GHz to 16 GHz
-20
-29
16 GHz to 20 GHz
-22
-33
Table 18
PNA or PNA-X Reflection Tracking Limits
Frequency
Ports 1 - 4
Ports 5 - 20
10 MHz to 50 MHz
1
1. Response from 10 MHz to 500 MHz is normal due to the PNA-X couplers in
comparison to the test set bridges. The bridges have more gain in the coupled
RF path.
-9
+16
50 MHz to 500 MHz
1
-10
-10
500 MHz to 3.2 GHz
-12
-18
3.2 GHz to 10 GHz
-17
-23
10 GHz to 16 GHz
-22
-29
16 GHz to 20 GHz
-27
-33
20 GHz to 24 GHz
-32
-39
24 GHz to 26.5 GHz
-37
-45
User's and Service Guide U3042-90012
57