IEEE-488.2 Common Commands
13
Keysight U2722A/U2723A Programmer’s Reference
119
*OPC/OPC?
Syntax
*OPC
The command is mainly used for program synchronization. It causes the
instrument to set the OPC bit (bit 0) of the Standard Event Status register when
the instrument has completed all pending operation sent before the
*OPC
command.
*OPC?
This query returns “1” to the output buffer at the completion of the current
operation.
*OPC?
command does not suspend processing of commands.
Remarks
Pending operation have completed when:
– all commands sent before
*OPC
, including paralleled commands, have
completed. Most commands are sequential and will be completed before the
next command is executed. Commands that affect the output voltage, current,
or state, relays, and trigger actions are executed in parallel with subsequent
commands.
*OPC
provides notification for the completion of the parallel
commands
– all triggered actions have completed
*OPC
does not prevent processing of subsequent commands, however, the OPC
bit will not be set until all pending operation have completed.
Returned Query Format
<NR1>
The command returns “1” to the output buffer.
See Also
Содержание u2722a
Страница 1: ...Keysight U2722A U2723A USB Modular Source Measure Units Programmer s Reference ...
Страница 7: ...Keysight U2722A U2723A Programmer s Reference 7 List of Figures Figure 1 Status system diagram 21 ...
Страница 8: ...8 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 10: ...10 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 50: ...8 SENSe Subsystem 50 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 102: ...10 STATus Subsystem 102 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...14 Error Messages 136 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...