13
IEEE-488.2 Common Commands
116
Keysight U2722A/U2723A Programmer’s Reference
*ESR?
Syntax
*ESR?
This query returns the value of the event register of the Standard Event Status
group. Once it is read, it will be cleared automatically. The event register is a
read-only register, which stores (latches) all standard events.
Remarks
– The following table lists the bit definitions for the Standard Event register.
– To be reported to the Standard Event register, the corresponding bits in the
event register must be enabled using the
– Once a bit is set, it will remain set until cleared by reading the event register or
the clear status (
) command.
Bit number
Decimal value
Definition
0 Operation Complete
1
All commands prior to and including *OPC have been executed.
1 Not Used
2
Always zero.
2 Query Error
4
The instrument tried to read the output buffer but it was empty. Or, a new
command line was received before a previous query has been read. Or, both
the input and output buffers are full.
3 Device-Dependent Error
8
A self-test or calibration error occurred. For a complete listing of the error
messages, refer to
4 Execution Error
16
An execution error occurred (an error in the –200 range has been generated.
5 Command Error
32
A command syntax error occurred (an error in the –100 range has been
generated).
6 Not Used
64
Always zero.
7 Power On
128
Power has been turned off and on since the last time the event register was
read or cleared.
Содержание u2722a
Страница 1: ...Keysight U2722A U2723A USB Modular Source Measure Units Programmer s Reference ...
Страница 7: ...Keysight U2722A U2723A Programmer s Reference 7 List of Figures Figure 1 Status system diagram 21 ...
Страница 8: ...8 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 10: ...10 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 50: ...8 SENSe Subsystem 50 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 102: ...10 STATus Subsystem 102 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...14 Error Messages 136 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...